Key result
Dual defibrillator use is linked to 39 reported events of device damage or temporary malfunction.
Why the study?
Dual defibrillator use is increasingly employed in elective and emergency settings, but device damage and malfunction associated with this practice require mapping and characterization.
Does dual defibrillator use cause device damage or malfunction?
Does dual defibrillator use cause device damage or malfunction?
Dual defibrillator use can cause device damage and malfunction that may remain occult until later testing, necessitating routine post-use device testing.
Answers are grounded in this paper and cite their sources.
Raises caution for dual defibrillator use amid device damage reports; hypothesis-generating and requires prospective validation.
Purpose To map and characterize device damage and malfunction associated with dual defibrillator use in defibrillation and cardioversion, which is increasingly employed in both elective and emergency settings, and to identify knowledge gaps relevant to practice and future research. Methods We conducted a scoping review in accordance with JBI methodology and PRISMA-ScR. PubMed, Embase, CINAHL, CENTRAL, and grey literature were searched without language or date restrictions. Regulatory adverse event databases were also searched. Eligible sources reported device damage or malfunction associated with two external defibrillators delivering shocks within one second or less of each other. Results Twenty-nine records from 2016 to 2026 were included: five published sources and 24 regulatory adverse event reports. Published sources comprised two randomised trials, one survey study, one case report, and one ad-hoc retrospective study. Across five published sources, eight events of lasting device damage and four temporary malfunctions were identified. In published studies, all reported damage events were reported in association with double simultaneous defibrillation or dual cardioversion. Across 24 regulatory reports, 11 records described device damage and 16 described malfunctions. Reported damage was often not recognized during the index event and was instead identified on post-use testing or later routine self-checks. Discussion Device damage and malfunction during dual defibrillator use are documented in both published and regulatory sources. Damage may remain occult until later testing, creating risk beyond the index case. Standardized prospective reporting and routine post-use device testing should be considered wherever dual defibrillator use is employed.
No takes yet. Share an insight, caveat, or question.
Purkarthofer et al. (2026) conducted a review in Dual defibrillator use in defibrillation and cardioversion (n=29). Dual defibrillator use was evaluated on Device damage and malfunction. Dual defibrillator use was associated with 19 events of lasting device damage and 20 temporary malfunctions across 29 published and regulatory records.