Key result
Mail-home Zio XT patches yield similar artifact rates to in-clinic application in pediatric patients.
Why the study?
During the COVID-19 pandemic, Zio XT patches were mailed directly to patients for self-application, prompting a comparison of data quality between in-clinic and mail-home applications in pediatrics.
Does mail-home application of continuous ambulatory ECG monitor patches provide equivalent tracing quality (percent artifact) compared to in-clinic application in pediatric patients?
Observational (n=284)
No
Does mail-home application of continuous ambulatory ECG monitor patches provide equivalent tracing quality (percent artifact) compared to in-clinic application in pediatric patients?
Mail-home application of ambulatory ECG patches in pediatric patients yields equivalent tracing quality to in-clinic application, supporting its use as a viable telemedicine alternative.
Answers are grounded in this paper and cite their sources.
Mail-home Zio application may be feasible in pediatrics; leaves open whether artifact rates match in-clinic use.
The Zio® (Zio) XT Patch is a 14-day continuous ambulatory ECG monitor. During the Covid-19 pandemic, Zios were mailed directly to patients for self-application. The purpose of this study was to compare the percent artifact, a marker for quality, of in-clinic (IC) to mail-home (MH) applications in a pediatric population. Methods: A single-center, IRB-approved study of patients 0-< 21 years of age with Zios were studied for wear and artifact time filtered out based on iRhythm’s proprietary algorithm. Results: In total, 284 Zios were randomly selected and analyzed for total wear time and artifact. Of these, 149 were IC prior to 12/31/2019 and 135 MH patches prescribed after 1/1/2020. No significant difference was found for percent artifact between the IC (7.8%) and MH (8.3%) group. Average IC wear-time was 127 hours compared to MH at 99 hours (p = 0.02). In conclusion, application of Zio patches outside of the pediatric cardiology clinic offers equivalent artifact, a marker of quality, as those applied in clinic and should be consideration as a viable alternative.
No takes yet. Share an insight, caveat, or question.
Khan et al. (2023) conducted an observational in Pediatric patients requiring ambulatory ECG monitoring (n=284). Mail-home application of Zio XT Patch vs. In-clinic application of Zio XT Patch was evaluated on Percent artifact (p=0.58). Mail-home application of Zio XT patches in pediatric patients yielded a similar percent artifact compared to in-clinic application, indicating equivalent quality.