Key result
ALFIS point-of-care cTnI assay correlates with the Atellica IM TnIH assay at ~0.85.
Why the study?
The authors sought to evaluate the analytical performance of ALFIS, a recently developed point-of-care chemiluminescence immunoassay for detecting cardiac troponin I.
The ALFIS point-of-care cardiac troponin I assay demonstrates acceptable imprecision but requires further optimization in analytical sensitivity, calibration, and EDTA interference management for clinical acute myocardial injury diagnosis.
Answers are grounded in this paper and cite their sources.
Cautions on EDTA interference with ALFIS cTnI assay; leaves open clinical reliability pending outcome validation.
Background:We evaluated the analytical performance of ALFIS (Boditech Med Inc., Korea), a recently developed point-of-care (POC) chemiluminescence immunoassay for the detection of cardiac troponin I (cTnI).Methods: Following the Clinical and Laboratory Standards Institute guidelines, we evaluated the imprecision, analytical sensitivity, method comparison, cross-reactivity, and interference of the assay kit.Results: The imprecision analysis yielded coefficients of variation (CVs) consistently below 10%.The assay's limit of blank and limit of detection values were 4 ng/L and 10 ng/L, respectively.The 10% CV was at 30 ng/L.The cTnI results between Atellica IM TnIH and ALFIS showed a Pearson correlation coefficient of 0.8478.The other cardiac biomarkers were not significantly cross-reactive, except for NT-proBNP mixed with low levels of cTnI.In the interference study, negative interference (>20% recovery) was evident with ethylenediaminetetraacetic acid at all tested concentrations of quality material. Conclusions:The ALFIS cTnI can be used as a POC diagnostic device for acute myocardial injury diagnosis, made possible by enhancing analytical performance parameters.
No takes yet. Share an insight, caveat, or question.
Kang et al. (2023) studied Myocardial injury (cardiac troponin I detection) (n=41). ALFIS cardiac troponin I assay vs. Atellica IM TnIH assay was evaluated on Method comparison (Pearson correlation coefficient) (Pearson correlation coefficient 0.8478, p=<0.0001). The ALFIS cardiac troponin I assay demonstrated a Pearson correlation coefficient of 0.8478 compared to the Atellica IM TnIH assay, with coefficients of variation consistently below 10%.