Key result
Origin-forced linear calibration improves apolipoprotein mass spectrometry stability with slope variation under ~12%.
Using origin-forced linear regression for mass spectrometric quantitation of apolipoproteins improves calibration stability and ensures long-term consistent quality.
Objectives Calibration is a cornerstone of clinical chemistry test results and maintaining an in-house LDT requires periodic renewal of calibrators. Here we present the introduction of a new set of native human serum calibrators, using different curve fitting approaches for long-term quantitation of nine apolipoproteins with traceability to current metrological traceability chains. Methods Five “old” native human serum calibrators (OC), have been used successfully for >6 years for quantitation of apolipoproteins by LC-MS. Five new calibrators (NC) were selected and value assigned by standard operating procedures. Calibration with OC used linear regression without constraining the intercept. In addition, regressions including the origin as a datapoint and forcing the origin were evaluated, and NC were recalculated using the OC under origin-forced conditions. Results The new calibrator set showed higher concentration ranges for all apos, except apoC-III. Including the origin as a datapoint in the linear regression improved stability of calibration lines (with reduction of slope variation and intercepts). Recalculation of the NCs and their subsequent application with origin forced, omitted intercepts of linear regression and in general reduced slope variations (<11.8 %CV, for all apos, except apo(a) with imprecision <9.1 % for all apos for two QC samples. Conclusions In a complex system like the mass spectrometric quantitation of multiple proteins, calibration with linear regression using origin forced appears to be a well-suited approach offering guaranteed long-term consistent quality and clinical use of the apolipoprotein test.
No takes yet. Share an insight, caveat, or question.
Smit et al. (2026) studied this question. New native human serum calibrators with origin-forced linear regression vs. Old calibrators with unconstrained linear regression was evaluated on Stability of calibration lines and slope variations. Calibration with linear regression using origin forced reduced slope variations (<11.8% CV) and improved stability for mass spectrometric quantitation of multiple apolipoproteins.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: