Key result
Cardiac devices in adult congenital heart disease carry high morbidity, with complications in ~33% of patients.
Why the study?
Permanent pacemakers and implantable cardioverter-defibrillators are often used to prevent arrhythmia-related morbidity or mortality in adult congenital heart disease, but they can also be associated with significant morbidity.
What are the rates of complications and inappropriate therapies for implantable anti-arrhythmic devices in adults with congenital heart disease?
Cohort (n=211)
No
What are the rates of complications and inappropriate therapies for implantable anti-arrhythmic devices in adults with congenital heart disease?
Adults with congenital heart disease face a uniquely high risk of complications and inappropriate shocks from pacemakers and ICDs, underscoring the need for specialized management.
Background: Brady- and tachyarrhythmias commonly complicate adult congenital heart disease (ACHD). Permanent pacemakers (PPMs) or implantable cardioverter-defibrillators (ICDs) are often utilised to prevent morbidity or mortality related to arrhythmia, but can also be associated with significant morbidity themselves. Methods: We analysed outcomes from patients in our comprehensive ACHD database who were seen at least twice since 2000 and once since 2018. Of 1953 ACHD patients, 134 had a PPM and 78 had an ICD (47 for primary and 31 for secondary prevention). Results: For PPM patients, 41% had a pacing percentage below 33%, 13% had 33-66%, and 46% had above 66%. One fifth required PPM upgrade, most to cardiac resynchronisation therapy, the rest to ICD. There were 33 appropriate ICD shocks in 15 patients (19%) and 34 inappropriate shocks in 13 patients (17%) over a median follow up of 4.6 years (IQR 0.9-8.3 years). Anti-tachycardia pacing was delivered appropriately for 28% of patients and inappropriately for 9%.Apart from inappropriate therapy, one third of PPM and ICD patients had other device-related complications. Acute PPM complications included lead dysfunction requiring revision (2%), pneumothorax (2%), pleural effusion (2%) and pocket infection (2%). ICDs were also acutely complicated by lead dysfunction (4%) as well as pocket hematoma (3%). The most common long-term complication overall was lead dysfunction, affecting one sixth of both PPM and ICD patients. Finally, the rate of device insertion increased significantly with disease severity. Conclusions: Anti-arrhythmic devices can be lifesaving in ACHD patients, but inappropriate therapy and device-related complications are very common.
No takes yet. Share an insight, caveat, or question.
Chami et al. (2023) conducted a cohort in Adult congenital heart disease (ACHD) (n=211). Permanent pacemakers (PPM) or implantable cardioverter-defibrillators (ICD) was evaluated on Device-related complications and inappropriate therapies. In adults with congenital heart disease, implantable anti-arrhythmic devices were associated with appropriate therapies but a high rate of device-related complications (33% of PPM and ICD patients) and inappropriate ICD shocks (17%).