Key result
Automated cuff BP devices show no significant difference from reference BP measurements in children.
Why the study?
Devices validated in adults may not be accurate in children, requiring separate evaluation, yet current pediatric guidelines still recommend confirming automated BP readings with auscultation.
Are automated cuff blood pressure devices accurate compared to reference standards in children aged 3-12 years?
Meta-Analysis (n=410)
Are automated cuff blood pressure devices accurate compared to reference standards in children aged 3-12 years?
While there is a severe shortage of properly validated automated BP devices for children, those that are successfully validated are accurate and can replace auscultatory devices for office hypertension diagnosis.
Answers are grounded in this paper and cite their sources.
Few validated automated devices exist for children; extends STRIDE BP database while leaving practice reliant on auscultatory confirmation.
BACKGROUND: Similar to adults, in children automated cuff devices are recommended for office, home, and ambulatory blood pressure (BP) measurement. However, a device validated in adults may not be accurate in children, and separate investigation is necessary. Moreover, current pediatric guidelines recommend hypertension detected by an automated device to be confirmed using auscultation. The evidence on validated automated BP devices for children (aged 3-12 years) is presented. METHODS: STRIDE BP (Science and Technology for Regional Innovation and Development in Europe-Blood Pressure; www.stridebp.org) performs periodic systematic PubMed searches for validation studies of automated BP devices. The data set of studies including children was analyzed. RESULTS: Fifty-one of 628 studies in the STRIDE BP database included children. Forty-one (80%) concluded that the test device passed the validation, yet due to protocol violations, STRIDE BP approved 22 of them (43%). These 22 studies validated 20 devices (14 on the market). Seven additional devices had measurement functions equivalent to STRIDE BP-approved devices (6 on the market). Meta-analysis of 20 successful studies including children and older individuals (N=1476) showed a pooled test-reference systolic BP difference of 0.6 mm Hg (95% CI, -0.2 to 1.4), and diastolic -0.3 (95% CI, -1.2 to 0.6). Meta-analysis of 7 studies including only children (N=410) showed a pooled difference of 1.0 (95% CI, -0.7 to 2.8)/-1.3 (95% CI, -2.9 to 0.4) mm Hg. CONCLUSIONS: There is a severe shortage of validated automated BP device models for children, with published studies often having methodological issues. Some devices are validated in children and, as in adults, can replace auscultatory devices for office hypertension diagnosis. There is an urgent need for more automated BP devices to be validated in children.
No takes yet. Share an insight, caveat, or question.
Stergiou et al. (2026) conducted a meta-analysis in Hypertension (n=410). Automated cuff blood pressure devices vs. Auscultation was evaluated on Pooled test-reference systolic BP difference (MD 1.0, 95% CI -0.7 to 2.8). Automated cuff BP devices showed a pooled test-reference systolic difference of 1.0 mm Hg and diastolic difference of -1.3 mm Hg in children.