Key result
Motion artifact removal circuit for integrated ECG sensors improves SNR by ~18 dB during abrupt movements.
Why the study?
Long-term real-time heart condition monitoring requires removing abrupt body motion artifacts without reliant digital post-processing, software algorithms, or off-chip components.
A novel analog circuit for ECG sensors effectively removes abrupt motion artifacts with very low power consumption, enabling better long-term real-time heart monitoring.
Answers are grounded in this paper and cite their sources.
May simplify ambulatory ECG monitoring without digital processing; leaves open clinical validation in patients.
This work describes an abrupt motion artifact removal circuit for integrated ECG sensor frontend suitable for long term real‐time heart condition monitoring. The proposed scheme removes artifacts due to abrupt body motions in the analog domain without using digital post processing or software algorithms, which does not require off‐chip components such as micro‐controllers or FPGAs. The key component of the proposed approach is the high‐pass and low‐pass filter that operates as an all‐pass filter for normal ECG inputs, whereas in the event of an abrupt motion, the high‐pass and low‐pass filter functions as a detection and delay element that detects motion induced high‐frequency components and blocks them at the input node of the ECG amplifier to prevent corruption such as output level saturation. The proposed circuit is realized using CMOS 180 nm technology (1‐poly, 6‐metal process) with a core area of 750 × 680 μm 2 and power consumption of 3.4 μW. Furthermore, the performance of the prototype device is verified with two subjects under a real‐life application setup, where the average SNR improvement across multiple trials showed 18.16 dB, 17.37 dB (for subject‐1) and 19.04 dB, 17.76 dB (for subject‐2) for removing artifacts due to two abrupt motions—standing up and answering a call that occurred during ECG recording. Overall, compared to prior art, the proposed circuit shows the lowest power consumption, comparable recovery time and area, which can lead to compact and low‐power ECG devices suitable for long‐term real‐time heart monitoring.
No takes yet. Share an insight, caveat, or question.
Nazari et al. (2026) studied heart condition monitoring (n=2). Abrupt motion artifact removal circuit vs. prior art was evaluated on SNR improvement. An abrupt motion artifact removal circuit for integrated ECG sensors achieved average SNR improvements of 17.37 to 19.04 dB across multiple trials in two subjects during abrupt motions.
Synapse has enriched 10 related papers. Consider them for comparative context: