Key result
Smaller-diameter bioprosthetic heart valves and thicker leaflets generated markedly higher fluttering frequencies in both experimental and computational models.
Why the study?
Bioprosthetic aortic valves can flutter in systole, which is linked to fatigue and failure, but the determinants of leaflet flutter in bioprosthetic heart valves have not been well characterized.
Smaller bioprosthetic heart valves and thicker leaflets generate higher fluttering frequencies, providing a potential mechanistic explanation for premature valve failure associated with prosthesis-patient mismatch.
May influence bioprosthetic valve design to reduce fluttering stress; leaves open clinical effects on human valve durability.
OBJECTIVE: Bioprosthetic heart valves (BHVs) are commonly used in surgical and percutaneous valve replacement. The durability of percutaneous valve replacement is unknown, but surgical valves have been shown to require reintervention after 10 to 15 years. Further, smaller-diameter surgical BHVs generally experience higher rates of prosthesis-patient mismatch, which leads to higher rates of failure. Bioprosthetic aortic valves can flutter in systole, and fluttering is associated with fatigue and failure in flexible structures. The determinants of flutter in BHVs have not been well characterized, despite their potential to influence durability. METHODS: We use an experimental pulse duplicator and a computational fluid-structure interaction model of this system to study the role of device geometry on BHV dynamics. The experimental system mimics physiological conditions, and the computational model enables precise control of leaflet biomechanics and flow conditions to isolate the effects of variations in BHV geometry on leaflet dynamics. RESULTS: Both experimental and computational models demonstrate that smaller-diameter BHVs yield markedly higher leaflet fluttering frequencies across a range of conditions. The computational model also predicts that fluttering frequency is directly related to leaflet thickness. A scaling model is introduced that rationalizes these findings. CONCLUSIONS: We systematically characterize the influence of BHV diameter and leaflet thickness on fluttering dynamics. Although this study does not determine how flutter influences device durability, increased flutter in smaller-diameter BHVs may explain how prosthesis-patient mismatch could induce BHV leaflet fatigue and failure. Ultimately, understanding the effects of device geometry on leaflet kinematics may lead to more durable valve replacements.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2020) studied Bioprosthetic heart valve fluttering (in vitro and computational model) (n=3). Smaller bioprosthetic heart valve diameter and thicker leaflets vs. Larger bioprosthetic heart valve diameter and thinner leaflets was evaluated on Leaflet fluttering frequency (proportionality coefficients of -7.79 Hz/mm (experimental) and -5.65 Hz/mm (computational) for diameter). Smaller-diameter bioprosthetic heart valves and thicker leaflets generated markedly higher fluttering frequencies in both experimental and computational models.