Key result
IC-ECG accurately identifies epicutaneous-cava catheter tip position in ~95% of neonates.
Why the study?
Evidence for intracavitary electrocardiography to locate 1-Fr epicutaneous-cava catheter tips in neonates remains limited.
Does intracavitary electrocardiography accurately locate 1-Fr epicutaneous-cava catheter tips in neonates compared to bedside echocardiography?
Observational (n=196)
No
Does intracavitary electrocardiography accurately locate 1-Fr epicutaneous-cava catheter tips in neonates compared to bedside echocardiography?
Intracavitary electrocardiography is a highly feasible and accurate method for guiding 1-Fr epicutaneous-cava catheter tip location in neonates.
Answers are grounded in this paper and cite their sources.
IC-ECG showed high accuracy for neonatal 1-Fr ECC tip location; leaves open need for prospective multicenter validation.
BACKGROUND: Evidence for intracavitary electrocardiography (IC-ECG) to locate 1-Fr epicutaneous-cava catheter (ECC) tips in neonates remains limited. METHODS: We retrospectively analyzed 196 neonates (2019-2025) undergoing 1-Fr ECC insertion with IC-ECG guidance using either a conventional monitor-defibrillator (DEF) or a dedicated neonatal interface (Pilot™). Bedside echocardiography performed at the end of the procedure (within 2 h) served as the reference standard. Endpoints were applicability, feasibility, accuracy, malposition, and early complications (⩽24 h). RESULTS: IC-ECG correctly identified catheter-tip position in 94.9% of cases under an intention-to-diagnose framework (186/196) and showed complete concordance with echocardiography in all evaluable catheters. Applicability was 100% (196/196; 95% CI 98.1-100.0). Feasibility (interpretable tracing) was 94.9% (186/196; 95% CI 90.8-97.5). Diagnostic accuracy was 94.9% under intention-to-diagnose (186/196; 95% CI 90.8-97.5) and 100% among interpretable tracings (186/186; 95% CI 98.0-100.0). In 10 cases, the tracing was neither readable nor interpretable, likely due to electrical interference. In three cases, the tracing was readable but the P-wave never reached a maximal amplitude throughout catheter navigation; ultrasound verification confirmed malposition. No clinically significant arrhythmias or occlusions occurred within 24 h. CONCLUSIONS: IC-ECG provides excellent diagnostic performance for 1-Fr ECC tip location in neonates. A flat P-wave should prompt ultrasound confirmation. Multicenter studies are warranted.
No takes yet. Share an insight, caveat, or question.
Gaetano et al. (2026) conducted an observational in Neonates requiring epicutaneous-cava catheter insertion (n=196). Intracavitary electrocardiography (IC-ECG) vs. Bedside echocardiography was evaluated on Diagnostic accuracy under intention-to-diagnose (95% CI 90.8-97.5). Intracavitary electrocardiography correctly identified 1-Fr epicutaneous-cava catheter tip position in 94.9% of neonates (95% CI 90.8-97.5) under an intention-to-diagnose framework.