Key result
The e-textile EASI ECG monitor produced signal quality comparable to a traditional Holter monitor during daily activities, indicating it is non-inferior to current clinical standards.
Why the study?
Cardiovascular disease places an increased burden on healthcare facilities, and e-textile-based cardiac monitoring offers a viable option for out-of-hospital cardiac rehabilitation programs.
Does an e-textile EASI ECG monitor provide comparable signal quality to a standard Holter monitor in a healthy adult?
Does an e-textile EASI ECG monitor provide comparable signal quality to a standard Holter monitor in a healthy adult?
Answers are grounded in this paper and cite their sources.
A novel e-textile EASI ECG monitor demonstrated non-inferior signal quality compared to a standard Holter monitor during daily activities in a healthy volunteer, suggesting its potential for ambulatory cardiac monitoring.
Background: With cardiovascular disease continuing to be the leading cause of death and the primary reason for hospitalization worldwide, there is an increased burden on healthcare facilities. Electronic-textile (e-textile)-based cardiac monitoring offers a viable option to allow cardiac rehabilitation programs to be conducted outside of the hospital. Objectives: This study aimed to determine whether signals produced by an e-textile ECG monitor with textile electrodes in an EASI configuration are of sufficient quality to be used for cardiac monitoring. Specific objectives were to investigate the effect of the textile electrode characteristics, placement, and condition on signal quality, and finally to compare results to a reference ECG obtained from a current clinical standard the Holter monitor. Methods: ECGs during different body movements (yawning, deep-breathing, coughing, sideways, and up movement) and activities of daily living (sitting, sitting/standing from a chair, and climbing stairs) were collected from a baseline standard of normal healthy adult male using a novel e-textile ECG and a reference Holter monitor. Each movement or activity was recorded for 5 min with 2-min intervals between each recording. Three different textile area electrodes (40, 60, and 70 mm 2 ) and electrode thicknesses (3, 5, and 10 mm) were considered in the experiment. The effect of electrode placement within the EASI configuration was also studied. Different signal quality parameters, including signal to noise ratio, approximate entropy, baseline power signal quality index, and QRS duration and QT intervals, were used to evaluate the accuracy and reliability of the textile-based ECG monitor. Results: The overall signal quality from the 70 mm 2 textile electrodes was higher compared to the smaller area electrodes. Results showed that the ECGs from 3 and 5 mm textile electrodes showed good quality. Regarding location, placing the “A” and “I” electrodes on the left and right anterior axillary points, respectively, showed higher signal quality compared to the standard EASI electrode placement. Wet textile electrodes showed better signal quality compared to their dry counterparts. When compared to the traditional Holter monitor, there was no significant difference in signal quality, which indicated textile monitoring was as good as current clinical standards (non-inferior). Conclusion: The e-textile EASI ECG monitor could be a viable option for real-time monitoring of cardiac activities. A clinical trial in a larger sample is recommended to validate the results in a clinical population.
No takes yet. Share an insight, caveat, or question.
Teferra et al. (2021) studied Healthy volunteer (n=1). e-textile EASI-based ECG monitor vs. Standard 3-lead Holter monitor was evaluated on Signal quality (ApEn, basSQI, SNR, QRS duration, QTc intervals). The e-textile EASI ECG monitor produced signal quality comparable to a traditional Holter monitor during daily activities, indicating it is non-inferior to current clinical standards.