Key result
Percutaneous sheath-in-sheath technique successfully retrieves a migrated PFO occluder from the abdominal aorta.
Why the study?
Migration of a cardiac occluder is a feared complication of endovascular therapy for interatrial defects.
Can a migrated PFO occluder in the abdominal aorta be successfully retrieved percutaneously using a small bore introducer?
Case Report (n=1)
Can a migrated PFO occluder in the abdominal aorta be successfully retrieved percutaneously using a small bore introducer?
This case demonstrates that percutaneous retrieval of a migrated PFO occluder from the abdominal aorta can be safely accomplished using a small bore (11 F) introducer and a sheath-in-sheath technique.
Demonstrates feasibility of sheath-in-sheath retrieval for migrated aortic PFO occluders; leaves open reproducibility and safety in larger cohorts.
Migration of a cardiac occluder is one of the most feared complications of endovascular therapy for interatrial defects. A 54-year-old female, after experiencing relapsing cryptogenic embolic cardiac ischaemia without coronary lesions, underwent percutaneous patent foramen ovale (PFO) occlusion with the placement of an Amplatzer Talisman. The next day, a scheduled check detected an early asymptomatic migration of the device to the abdominal aorta. A percutaneous attempt involving interventional radiologists was chosen. Through the common femoral artery, the right atrial screw was targeted and snared. The PFO occluder was thus easily unsheathed utilizing a sheath-in-sheath technique. Endovascular retrieval was successfully performed through an 11 F introducer. This case reminds us that migration is a potential complication of any atrial septal endovascular occlusion and shows that retrieval may be percutaneously accomplished with a small bore introducer.
No takes yet. Share an insight, caveat, or question.
Napoli et al. (2026) conducted a case report in Migrated patent foramen ovale (PFO) occluder (n=1). Percutaneous retrieval using a sheath-in-sheath technique through an 11 F introducer was evaluated on Successful endovascular retrieval. Percutaneous retrieval of a migrated PFO occluder from the abdominal aorta was successfully accomplished using a sheath-in-sheath technique through a small bore 11 F introducer.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: