Key result
Transcatheter retained super-stiff guide wire technique successfully retrieves a dislodged ASD occluder from the aorta.
Why the study?
Device embolisation is one of the most feared early complications of percutaneous atrial septal defect closure.
Can the 'retained super-stiff guide wire' technique safely retrieve an embolised atrial septal defect occluder from the descending aorta?
Case Report (n=1)
Can the 'retained super-stiff guide wire' technique safely retrieve an embolised atrial septal defect occluder from the descending aorta?
The 'retained super-stiff guide wire' technique allows for safe trans-arterial retrieval of embolised cardiac devices from the aorta.
Answers are grounded in this paper and cite their sources.
Supports percutaneous retrieval of embolized ASD occluders without aortic trauma in select cases; hypothesis-generating and should not yet change practice.
Device embolisation is one of the most feared early complications of percutaneous atrial septal defect closure. We report a 40-year-old woman in whom a 12-mm atrial septal defect occluder was accidentally displaced into the descending aorta 24 hours after an apparently successful procedure. The dislodged device was retrieved percutaneously through a 12-F long sheath that was continuously railroaded over a super-stiff guide wire. This "retained super-stiff guide wire" technique secured sheath stability, avoided aortic wall trauma, and allowed the occluder to be slipped into the sheath without change of its orientation. A larger 16-mm small-waist-big-edge device was subsequently implanted uneventfully. The technical details and advantages of keeping the sheath constantly over a super-stiff guide wire are highlighted to facilitate safe trans-arterial retrieval of embolised cardiac devices.
No takes yet. Share an insight, caveat, or question.
Xiao et al. (2026) conducted a case report in Device embolisation of atrial septal defect occluder (n=1). Transcatheter retrieval using the 'retained super-stiff guide wire' technique was evaluated on Successful retrieval of the dislodged device. Transcatheter retrieval using a 'retained super-stiff guide wire' technique successfully removed a fully dislodged atrial septal defect occluder from the descending aorta.
Synapse has enriched 10 related papers. Consider them for comparative context: