Key result
A normalized index of hemolysis expressed in milligrams is proposed as the most useful formula for standardizing the assessment of hemolysis in rotary blood pumps.
Why the study?
What is the most appropriate formula for standardizing the index of hemolysis in the in vitro evaluation of rotary blood pumps?
What is the most appropriate formula for standardizing the index of hemolysis in the in vitro evaluation of rotary blood pumps?
Answers are grounded in this paper and cite their sources.
Proposes standardizing the evaluation of blood pump hemolysis using a normalized index expressed in milligrams to improve comparability across devices.
Hemolysis is one of the most important performance parameters of blood pumps. However, comparative in vitro evaluation of the reported hemolysis effect is difficult owing to the lack of uniformity in the test methods used. Currently, three types of formulas are generally accepted and widely used for the index of hemolysis: the traditionally used index of hemolysis (defined as grams of plasma free hemoglobin released per 100 L of blood pumped); the normalized index of hemolysis (normalized by hematocrit); and the modified index of hemolysis (taking into account not only hematocrit but also hemoglobin). In addition, the tested blood conditions are often not indicated in the reports. To address this confusing situation, all three indices of hemolysis were routinely derived in our laboratory. To avoid further confusion, the tested blood conditions and test loop were defined in each study in our laboratory. If we limit the acceptable range of hemoglobin in human blood, the normalized index of hemolysis is sufficient. Furthermore, using milligrams as the unit of expression would be easier and reasonable for comparing the less hemolytic blood pumps, such as the centrifugal pumps that are currently available clinically. We would like to propose this value of a normalized index of hemolysis in milligrams as the most useful formula for the index of hemolysis in the assessment of various types of rotary blood pumps.
No takes yet. Share an insight, caveat, or question.
Naito et al. (1994) studied Hemolysis in blood pumps. Standardized index of hemolysis was evaluated. A normalized index of hemolysis expressed in milligrams is proposed as the most useful formula for standardizing the assessment of hemolysis in rotary blood pumps.