Key result
High-power electric vehicle chargers cause no electromagnetic interference in patients with CIEDs.
Why the study?
New high-power electric vehicle charging stations have the potential to induce electromagnetic interference in CIEDs, but their safety had not been evaluated.
Does the use of high-power electric vehicle chargers cause clinically relevant electromagnetic interference in patients with cardiac implantable electronic devices?
Observational (n=130)
Does the use of high-power electric vehicle chargers cause clinically relevant electromagnetic interference in patients with cardiac implantable electronic devices?
High-power chargers for battery electric vehicles appear safe for patients with cardiac implantable electronic devices, causing no detectable electromagnetic interference.
High-power EV chargers appear safe for CIED patients; leaves open need for larger prospective confirmation.
AIMS: Battery electric vehicle (BEV) sales and use are rapidly expanding. Battery electric vehicles, along with their charging stations, are a potential source of electromagnetic interference (EMI) for patients with cardiac implantable electronic devices (CIEDs). The new 'high-power' charging stations have the potential to create strong electromagnetic fields and induce EMI in CIEDs, and their safety has not been evaluated. METHODS AND RESULTS: A total of 130 CIED patients performed 561 charges of four BEVs and a test vehicle (350 kW charge capacity) using high-power charging stations under continuous 6-lead electrocardiogram monitoring. The charging cable was placed directly over the CIED, and devices were programmed to maximize the chance of EMI detection. Cardiac implantable electronic devices were re-interrogated after patients charged all BEVs and the test vehicle for evidence of EMI. There were no incidences of EMI, specifically no over-sensing, pacing inhibition, inappropriate tachycardia detection, mode switching, or spontaneous reprogramming. The risk of EMI on a patient-based analysis is 0/130 [95% confidence interval (CI) 0%-2%], and the risk of EMI on a charge-based analysis is 0/561 (95% CI 0%-0.6%). The effective magnetic field along the charging cable was 38.65 µT and at the charging station was 77.9 µT. CONCLUSIONS: The use of electric cars with high-power chargers by patients with cardiac devices appears to be safe with no evidence of clinically relevant EMI. Reasonable caution, by minimizing the time spent in close proximity with the charging cables, is still advised as the occurrence of very rare events cannot be excluded from our results.
No takes yet. Share an insight, caveat, or question.
Lennerz et al. (2023) conducted an observational in Cardiac implantable electronic devices (CIEDs) (n=130). High-power chargers for battery electric vehicles was evaluated on Incidence of electromagnetic interference (EMI) (95% CI 0%-2%). The use of high-power chargers for electric vehicles caused no episodes of clinically relevant electromagnetic interference in patients with cardiac implantable electronic devices.