A novel silicon microdosimeter with a 3D cylindrical “dartboard” geometry has been developed at the National Center of Microelectronics (IMB-CNM) in Barcelona, Spain. The detector features concentric 3D electrodes etched into a 20 μ m -thick sensitive silicon volume. Its performance has been characterized using the ion beam–induced charge (IBIC) technique, and its radiation hardness has been investigated in view of potential applications in proton therapy. The central region was irradiated with 1.17 MeV protons up to a cumulative fluence of 1 . 7 × 1 0 12 cm − 2 , with a bias voltage of 5 V. Spatially resolved charge collection efficiency (CCE) maps and pulse height spectra were analyzed as a function of the accumulated fluence. No significant degradation was observed for fluences below 6 × 1 0 10 cm − 2 . At higher fluences, however, progressive CCE reduction occurs, accompanied by peak broadening, spectral redistribution towards lower pulse height and contraction of the effective high-efficiency sensor region. This degradation is more pronounced at the periphery of the sensitive volume, which is consistent with the radial electric field distribution inherent to the 3D geometry.