Key points are not available for this paper at this time.
Despite the significant demand, accurate determination of the permanganate index in high-chloride bodies of water remains extremely limited. This study introduces an iodine-thiosulfate method that effectively eliminates the interference of chloride ions in the permanganate index determination. In this method, organic and inorganic substances in a water sample are oxidized with KMnO 4 in an alkaline medium to mitigate chloride ion interference. To further minimize this interference during back-titration, the remaining KMnO 4 is reduced using KI under adjusted acidic conditions. The released iodine is then titrated with standardized Na 2 S 2 O 3 until the starch–iodine complex’s blue–black color disappears. The iodine-thiosulfate method demonstrated a method detection limit of 0.4 mg L –1 ( n = 7) and a corresponding interlaboratory quantification limit of 1.6 mg L –1 in water samples with chloride concentrations of 5000 mg L –1 . The method’s precision and accuracy ranged from 1.4% to 6.7% and −1.0% to 5.0%, respectively. The relative error in permanganate index determination remained below 20% even at chloride concentrations up to 60 g L –1, whereas the conventional oxalate-permanganate method exceeded a 20% relative error once the chloride concentration reached 10 g L –1 or higher. The sufficiently low detection limit along with excellent repeatability (intralaboratory precision), reproducibility (interlaboratory precision), and accuracy confirms its practical feasibility for routine analysis.
Wang et al. (Mon,) studied this question.