Key result
Fluid-filled catheters overestimate LV and aortic systolic pressures by ~7 mmHg versus solid-state catheters.
Why the study?
To compare left ventricular and aortic pressures obtained using fluid-filled versus high-fidelity solid-state pressure catheters during left heart catheterization.
Do fluid-filled pressure catheters provide different left ventricular and aortic pressure measurements compared to solid-state pressure catheters in subjects undergoing left heart catheterization?
Do fluid-filled pressure catheters provide different left ventricular and aortic pressure measurements compared to solid-state pressure catheters in subjects undergoing left heart catheterization?
Fluid-filled catheters significantly overestimate true systolic pressures in the left ventricle and aorta compared to high-fidelity solid-state catheters during left heart catheterization.
Fluid-filled catheters may overestimate LVSP; leaves open standardization of invasive hemodynamic measurements.
Objective: To compare left ventricular (LV) and aortic (AO) pressures obtained using fluid‐filled and high‐fidelity solid‐state pressure catheters in subjects undergoing left heart catheterization. Materials and Methods: Twenty subjects scheduled for a left heart catheterization were enrolled and 18 subjects completed this IRB‐approved study. LV and AO pressures were obtained using fluid‐filled pressure catheter (standard‐of‐care) and high‐fidelity solid‐state pressure catheter synchronously. Pressure tracings were analyzed to measure LV systolic (LVSP), LV minimum‐diastolic (LVMDP), LV end‐diastolic (LVEDP), AO systolic (AOSP), and AO diastolic (AODP) pressures. Isovolumic contraction and relaxation rates (peak ± dp/dt) were derived from the pressure waveforms. Repeated measures of variance, post hoc tests with Bonferroni corrections, and Bland–Altman plots were used for comparisons. Results: A significant main effect of the pressure catheter was noted for LVSP, LVMDP, and AOSP ( p ≤ 0.025). The LVSP and AOSP measured with fluid‐filled pressure catheters were higher by 6.6 ± 6.9 mmHg and 4.6 ± 5.2 mmHg in comparison to solid‐state pressure catheter. In contrast, the LVMDP measurements were 3.5 ± 5.7 mmHg lower than the solid‐state pressure catheter measurements. The isovolumic contraction (66.4 ± 116.0 mmHg/s) and relaxation rates (60.5 ± 113.5 mmHg/s) were not significantly different between catheter systems after Bonferroni corrections for multiple comparisons ( p ≥ 0.06). The Bland–Altman analysis revealed a bias ranging from 0.8 to 6.6 mmHg. Conclusions: Differences in LVSP, LVMDP, and AOSP were noted between the catheter systems but not for other pressure values and contraction/relaxation rates. Fluid‐filled catheters overestimated true systolic pressures in the left ventricle and aorta. Trial Registration: ClinicalTrials.gov identifier: NCT03245255
No takes yet. Share an insight, caveat, or question.
Machado et al. (2025) studied Scheduled for left heart catheterization (n=18). Fluid-filled pressure catheter vs. High-fidelity solid-state pressure catheter was evaluated on Left ventricular systolic pressure (LVSP) (MD 6.6 mmHg, p=≤0.025). Fluid-filled pressure catheters overestimated left ventricular and aortic systolic pressures compared to solid-state catheters (LVSP higher by 6.6 ± 6.9 mmHg, p≤0.025).