A new hydrophilic interaction ultra-performance LC method was established for the whole blood measurement of L-ergothioneine. Chromatographic separation was achieved in a fairly short time, less than 4 min, on a 100 × 2.1 mm Acquity UPLC BEH HILIC 1.7 μm column with a mobile phase consisting of a mixture of 100 mmol/L ammonium acetate/ACN/water (5:85:10, v/v/v) that flowed isocratically at 0.250 mL/min. The LOD and the limit of quantification were 3.85 and 11.67 μmol/L, respectively. The method exhibited linearity in a concentration range of 15.63-1000 μmol/L (R(2) > 0.999). Mean recovery was 96.34% whereas intraassay and interassay precision were 1.52 and 1.82% RSD, respectively. On the whole, the developed method is simple, fast, precise, accurate, and sensitive and may be useful for routine analyses.
No takes yet. Share an insight, caveat, or question.
Sotgia et al. (2012) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: