Key result
Mid-infrared laser angioplasty yields ~87% standalone procedural success, but 6-month restenosis remains high.
Observational (n=1,862)
Yes
Answers are grounded in this paper and cite their sources.
BACKGROUND AND OBJECTIVE: The solid-state, mid-infrared holmium:YAG laser (2.1 microm wavelength) is a relatively new percutaneous device that has recently been evaluated in a multicenter study. Because of its unique wavelength and photoacoustic effects on atherosclerotic plaques, this laser may be useful in treatment of symptomatic patients with coronary artery disease. This study sought to evaluate the safety and efficacy of mid-infrared laser angioplasty in the treatment of coronary artery lesions. PATIENTS AND METHODS: Laser angioplasty was performed on 2,038 atherosclerotic lesions in 1,862 consecutive patients with a mean age of 61 +/- 11 years. Clinical indications included unstable angina (69%), stable angina (20%), acute infarction (6%), and positive exercise test (5%). Complex lesion morphology included eccentricity (62%), thrombus (30%), total occlusion (27%), long lesions (14%), and saphenous vein grafts (11%). RESULTS: This laser catheter alone successfully reduced stenosis (>20%) in 87% of lesions. With adjunct balloon angioplasty, 93% procedural success was achieved. The presence of thrombus within the target lesion was a predictor of procedural success (OR = 2.0 [95% confidence interval 2.0, 4.0], P = .04). Bifurcation lesions (OR = 0.5 [95% confidence interval 0.2, 1.0], P = .05) and severe tortuosity of the treated vessel (OR = 0.4 [95% confidence interval 0.2, 0.9], P = .02) were identified as significant predictors of decreased laser success. Calcium within the lesion was associated with reduced procedural success (OR = 0.57 [95% confidence interval 0.34, 0.97], P = .03), and calcified lesions required significantly more energy pulses than noncalcified lesions (119 +/- 91 pulses vs. 101 +/- 86 pulses, respectively, P = .0002). Complications included in-hospital bypass surgery 2.5%, Q-wave myocardial infarction 1.2%, and death 0.8%. Perforation occurred in 2.2% of patients; major dissection in 5.8% of patients, and spasm in 12% of patients. No predictor of major complications was identified. Six-month angiographic restenosis was documented in 54% of patients, and clinical restenosis occurred in 34% of patients. CONCLUSION: Mid-infrared laser has a safety profile similar to that of other debulking devices. This laser may be useful in select patients presenting with acute ischemic syndromes associated with intracoronary thrombus; however, like other coronary lasers, it is limited by the need for adjunctive balloon angioplasty and/or stenting to achieve adequate final luminal diameter. No beneficial effects on reducing 6-month restenosis rates were observed.
No takes yet. Share an insight, caveat, or question.
Topaz et al. (1998) conducted an observational in Coronary artery disease (n=1,862). Mid-infrared holmium:YAG laser angioplasty was evaluated on Procedural success (reduction of stenosis >20%) with laser catheter alone. Mid-infrared laser angioplasty achieved 87% standalone and 93% overall procedural success with balloon adjunct, though 6-month angiographic restenosis remained high at 54%.
Synapse has enriched 10 related papers. Consider them for comparative context: