Answers are grounded in this paper and cite their sources.
Examines bioheat transfer during cryosurgery, suggesting new insights for surgical applications.
The transmission line matrix (TLM) method is developed to efficiently solve the bioheat transfer with phase change in freeze-thaw cryoablation. This new design of three dimensional TLM model with an automatic time stepping is based on hyperbolic heat transfer model. It is able to consider many factors such as temperature-dependant thermophysical properties during phase change, blood perfusion and latent heat release/absorption. Thermal analysis of heat transfer in the cryotreated tissue is detailed; with special emphases on lethal temperature isotherms with respect to the cooling rates and duration of freeze-thaw cycle. The basic numerical model is validated through available experimental data on skeletal muscle of rabbit hind-limbs.
No takes yet. Share an insight, caveat, or question.
Bellil et al. (2018) studied this question.