Key result
In a porcine model, radiofrequency ablation using a perfusion electrode achieved a significantly larger coagulation volume (16.9 cm3) than internally cooled (7.2 cm3) or multitined electrodes (7.5 cm3) (p<0.05).
Why the study?
Does RF ablation with a perfusion electrode increase coagulation volume compared to internally cooled or multitined expandable electrodes in porcine lungs?
Does RF ablation with a perfusion electrode increase coagulation volume compared to internally cooled or multitined expandable electrodes in porcine lungs?
Absolute Event Rate: 16.9% vs 7.2%
p-value: p=<0.05
In a porcine lung model, RF ablation using a perfusion electrode achieved a significantly larger coagulation volume compared to internally cooled or multitined expandable electrodes.
Preclinical electrode comparisons in porcine lung remain hypothesis-generating; leaves open optimal device selection for clinical ablation.
The purpose of this study was to compare the amounts of in vivo coagulation obtained by radiofrequency (RF) ablation in porcine lung, using three types of electrodes. 15 in vivo ablation procedures were performed in the lungs of five pigs using three kinds of currently available RF devices under CT guidance. After placing an electrode in the lung, three ablation zones were created at each of three different regimens: Group A: RF ablation with an internally cooled electrode; Group B: RF ablation with a perfusion electrode, with instillation of 0.9% NaCl solution at a rate of 1.5 ml min(-1); Group C: RF ablation with a multitined expandable electrode. According to the manufacturer's recommendations, RF application times were 12 min in group A and 20 min in group B. In group C, RF energy was delivered for 7 min after a mean temperature of 110 degrees C was reached at 5 cm deployment. 36 min after the procedures, contrast-enhanced CT scans were obtained to evaluate the volume of zone of coagulation, and lungs were harvested for gross measurements. After macroscopic and histopathological analyses of 5 mm-thick lung sections, diameters, volumes and variation coefficients of regions of central coagulation were assessed. During RF ablation, the perfusion electrode allowed a larger energy delivery than the internally cooled or the multitined expandable electrodes, i.e. 33.6+/-4.7 kJ in group A, 40.0+/-8.2 kJ in group B and 23.5+/-6.1 kJ in group C (p<0.05). On gross observation, the cut surface of the gross specimen containing RF-induced coagulation showed that the ablated tissue appeared to be a central, firm, dark-brown area surrounded by an irregular outer margin (approximately 3-10 mm thick) of bright red tissue. In vivo studies showed that RF ablation using the perfusion electrode achieved larger coagulation volume than RF ablation using the other electrodes (p<0.05): 7.2+/-4.1 cm3 in group A; 16.9+/-5.5 cm3 in group B; 7.5+/-3.3 cm3 in group C. The corresponding variation coefficients were 0.55, 0.31, and 0.45, respectively. Our study shows that RF ablation using a perfusion electrode achieves a larger coagulation volume with an irregular margin than RF ablation using internally cooled or multitined expandable electrodes in the porcine lung.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2006) studied Normal porcine lung (n=5). Radiofrequency ablation with a perfusion electrode vs. Internally cooled electrode and multitined expandable electrode was evaluated on Volume of zone of coagulation (p=<0.05). In a porcine model, radiofrequency ablation using a perfusion electrode achieved a significantly larger coagulation volume (16.9 cm3) than internally cooled (7.2 cm3) or multitined electrodes (7.5 cm3) (p<0.05).