Key result
Following TMVR with LAMPOON, neo-LVOT and skirt neo-LVOT areas varied substantially throughout systole (64 ± 27% and 25 ± 14% change), with smaller areas in subjects with LVOT obstruction.
Why the study?
To characterize the dynamic nature of LVOT geometry and flow rate in patients after TMVR with LAMPOON to help guide future patient selection.
Observational (n=19)
LVOT geometry is highly dynamic following TMVR with LAMPOON, suggesting that phase-specific threshold values for skirt neo-LVOT area should be used to guide patient selection and avoid LVOT obstruction.
Answers are grounded in this paper and cite their sources.
Dynamic LVOT characterization post-LAMPOON TMVR may refine selection; leaves open prospective validation in larger cohorts.
AIMS: To characterize the dynamic nature of the left ventricular outflow tract (LVOT) geometry and flow rate in patients following transcatheter mitral valve replacement (TMVR) with anterior leaflet laceration (LAMPOON) and derive insights to help guide future patient selection. METHODS AND RESULTS: Time-resolved LVOT geometry and haemodynamics were analysed with post-procedure computed tomography and echocardiography in subjects (N = 19) from the LAMPOON investigational device exemption trial. A novel post hoc definition for LVOT obstruction was employed to account for systolic flow rate and quality of life improvement [obstruction was defined as LVOT gradient >30 mmHg or LVOT effective orifice area (EOA) ≤1.15 cm2]. The neo-LVOT and skirt neo-LVOT were observed to vary substantially in area throughout systole (64 ± 27% and 25 ± 14% change in area, respectively). The peak systolic flow rate occurred most commonly just prior to mid-systole, while minimum neo-LVOT (and skirt neo-LVOT) area occurred most commonly in early-diastole. Subjects with LVOT obstruction (n = 5) had smaller skirt neo-LVOT values across systole. Optimal thresholds for skirt neo-LVOT area were phase-specific (260, 210, 200, and 180 mm2 for early-systole, peak flow, mid-systole, and end-systole, respectively). CONCLUSION: The LVOT geometry and flow rate exhibit dynamic characteristics following TMVR with LAMPOON. Subjects with LVOT obstruction had smaller skirt neo-LVOT areas across systole. The authors recommend the use of phase-specific threshold values for skirt neo-LVOT area to guide future patient selection for this procedure. LVOT EOA is a 'flow-independent' metric which has the potential to aid in characterizing LVOT obstruction severity.
No takes yet. Share an insight, caveat, or question.
Kohli et al. (2021) conducted an observational in Patients following transcatheter mitral valve replacement (TMVR) with anterior leaflet laceration (LAMPOON) (n=19). Transcatheter mitral valve replacement (TMVR) with LAMPOON was evaluated on Time-resolved LVOT geometry and haemodynamics (neo-LVOT and skirt neo-LVOT area variation). Following TMVR with LAMPOON, neo-LVOT and skirt neo-LVOT areas varied substantially throughout systole (64 ± 27% and 25 ± 14% change), with smaller areas in subjects with LVOT obstruction.