Key result
The CALERA Research wearable temperature sensor demonstrated excellent reliability and substantial agreement with a rectal probe for determining the circadian phase of core body temperature.
Why the study?
To evaluate whether a patch-type wearable temperature sensor could determine the circadian phase of core body temperature similarly to a rectal probe in a real-world setting.
Does a patch-type wearable temperature sensor accurately determine the circadian phase of core body temperature compared to a rectal probe?
Observational (n=16)
No
Does a patch-type wearable temperature sensor accurately determine the circadian phase of core body temperature compared to a rectal probe?
The CALERA Research wearable temperature sensor provides a reliable and acceptable surrogate for rectal probes in estimating the circadian phase of core body temperature.
Answers are grounded in this paper and cite their sources.
May enable noninvasive CBT phase estimation in real-world settings; leaves open larger validation before clinical adoption.
To evaluate whether a patch-type wearable temperature sensor (CALERA Research) could determine the circadian phase of core body temperature (CBT) in a manner like a rectal probe. Sixteen participants (27 ± 11 years, 8 males and 8 females) wore an actigraph and CALERA Research sensor on the chest region for 3–5 days in a real-world setting. Simultaneous rectal temperature measurements were performed during the nocturnal sleep period. The midpoints of the nocturnal decrease in CBT (CBTtrough) were used as the circadian phase marker. We analyzed 60 pairs of CBTtrough. The reliability and agreement of the CBTtrough from the two devices were analyzed using the intraclass correlation coefficient (ICC) and the concordance correlation coefficient (CCC). The Bland-Altman analysis was used to quantify the limit of agreement of CBTtrough between the devices. The ICC of 0.96 (95%CI: 0.93–0.98) and CCC of 0.96 (95%CI: 0.93–0.97) values indicated excellent reliability and substantial agreement, respectively. The mean bias was 0.16 hours (95%LoA: -0.76–1.07 hours). The mean CBTtrough comparison was 5.9 ± 1.6 hours in the CALERA Research sensor and 5.8 ± 1.7 hours in the rectal probe. The difference in the CBTtrough between the two devices was about ± 1.0 hour which would be an acceptable range for determining the CBTtrough. We suggest that the CALERA Research sensor could be a useful tool for reasonably estimating the circadian phase of CBTtrough and providing a surrogate for a rectal probe.
No takes yet. Share an insight, caveat, or question.
Kubota et al. (2025) conducted an observational in Healthy adults (circadian rhythm assessment) (n=16). CALERA Research wearable temperature sensor vs. Rectal temperature probe was evaluated on Reliability and agreement of the midpoint of nocturnal decrease in core body temperature (CBT trough) (ICC 0.96, 95% CI 0.93-0.98). The CALERA Research wearable temperature sensor demonstrated excellent reliability and substantial agreement with a rectal probe for determining the circadian phase of core body temperature.
Synapse has enriched 10 related papers. Consider them for comparative context: