Key result
The SelectSecure Lead System resulted in 0% vein occlusion or blood flow alteration in 20 pediatric patients after a median follow-up of 19 months.
Why the study?
Does selective-site pacing using the SelectSecure Lead System cause subclavian vein occlusion in paediatric patients with atrioventricular block?
Observational (n=20)
Does selective-site pacing using the SelectSecure Lead System cause subclavian vein occlusion in paediatric patients with atrioventricular block?
Answers are grounded in this paper and cite their sources.
The SelectSecure Lead System appears safe for permanent intracardiac pacing in pediatric patients without increasing the risk of venous thrombosis over a median 19 months of follow-up.
AIMS: Few data are available about selective-site pacing in children. A 4.1 Fr lead is now available: it has to be introduced through an 8.4 Fr steerable catheter, and it raises concerns as to the long-term patency of the subclavian vein. The aim of this study was to systematically evaluate the patency of the subclavian vein in children who had selective-site pacing by the SelectSecure Lead System (SSLS) for atrioventricular block and look for possible local early and late complications. METHODS AND RESULTS: From June 2006 to January 2010, we implanted 20 patients. Mean age at implantation was 9 years (range 3-16), mean weight 31 kg (range 13-57). All patients received a dual-chamber pacemaker (PM). The patency of the subclavian vein was evaluated by ultrasound 1, 3, and every 6 months after the PM implantation. We also looked for oedema, swelling, changes in temperature or colour of the skin, and visible superficial circulation. Forty leads were implanted in 20 patients. After a median follow-up of 19 months (range 6-44), no clinical signs or symptoms of vein occlusion were detected in the whole population. None of the patients had an alteration of blood flow at Doppler examination. CONCLUSION: The SSLS is a promising and safe system for permanent intracardiac pacing in paediatric patients. We show that it does not increase the risk of venous thrombosis. Until a venography study with a larger population and longer follow-up regarding venous patency and overall complications using SSLS will be performed, our results although promising have to be considered preliminary data.
No takes yet. Share an insight, caveat, or question.
Gabbarini et al. (2010) conducted an observational in Atrioventricular block (n=20). SelectSecure Lead System (SSLS) was evaluated on Patency of the subclavian vein (clinical signs of vein occlusion or alteration of blood flow at Doppler examination). The SelectSecure Lead System resulted in 0% vein occlusion or blood flow alteration in 20 pediatric patients after a median follow-up of 19 months.
Synapse has enriched 10 related papers. Consider them for comparative context: