Answers are grounded in this paper and cite their sources.
We present a compact sensor for carbon monoxide (CO) in air and exhaled breath based on a room temperature interband cascade laser (ICL) operating at 4.69 µm, a low-volume circular multipass cell and wavelength modulation absorption spectroscopy. A fringe-limited (1σ) sensitivity of 6.5 × 10-8 cm-1Hz-1/2 and a detection limit of 9 ± 5 ppbv at 0.07 s acquisition time are achieved, which constitutes a 25-fold improvement compared to direct absorption spectroscopy. Integration over 10 s increases the precision to 0.6 ppbv. The setup also allows measuring the stable isotope 13CO in breath. We demonstrate quantification of indoor air CO and real-time detection of CO expirograms from healthy non-smokers and a healthy smoker before and after smoking. Isotope ratio analysis indicates depletion of 13CO in breath compared to natural abundance.
No takes yet. Share an insight, caveat, or question.
Ghorbani et al. (2017) studied this question.