Key result
The Tempo temporary pacing lead was successfully positioned and achieved pacing in 98.1% of patients undergoing structural heart procedures, with no clinically significant perforations or effusions.
Why the study?
Despite increasing use of temporary cardiac pacing in structural heart procedures, temporary pacing leads have not significantly improved, prompting evaluation of the safety and efficacy of the novel active-fixation Tempo Temporary Pacing Lead.
Is the Tempo active fixation temporary pacing lead safe and effective for temporary cardiac pacing during structural heart procedures?
Observational (n=269)
Yes
Is the Tempo active fixation temporary pacing lead safe and effective for temporary cardiac pacing during structural heart procedures?
The Tempo active fixation temporary pacing lead is safe and effective for structural heart procedures, providing stable peri- and postprocedural pacing while allowing for patient mobilization.
May support safer temporary pacing in structural procedures; leaves open need for randomized comparisons versus conventional leads.
OBJECTIVES: This multicenter retrospective study of the initial U.S. experience evaluated the safety and efficacy of temporary cardiac pacing with the Tempo® Temporary Pacing Lead. BACKGROUND: Despite increasing use of temporary cardiac pacing with the rapid growth of structural heart procedures, temporary pacing leads have not significantly improved. The Tempo lead is a new temporary pacing lead with a soft tip intended to minimize the risk of perforation and a novel active fixation mechanism designed to enhance lead stability. METHODS: Data from 269 consecutive structural heart procedures were collected. Outcomes included device safety (absence of clinically significant cardiac perforation, new pericardial effusion, or sustained ventricular arrhythmia) and efficacy (clinically acceptable pacing thresholds with successful pace capture throughout the index procedure). Postprocedure practices and sustained lead performance were also analyzed. RESULTS: The Tempo lead was successfully positioned in the right ventricle and achieved pacing in 264 of 269 patients (98.1%). Two patients (0.8%) experienced loss of pace capture. Procedural mean pace capture threshold (PCT) was 0.7 ± 0.8 mA. There were no clinically significant perforations, pericardial effusions, or sustained device-related arrhythmias. The Tempo lead was left in place postprocedure in 189 patients (71.6%) for mean duration of 43.3 ± 0.7 hr (range 2.5-221.3 hr) with final PCT of 0.84 ± 1.04 mA (n = 80). Of these patients, 84.1% mobilized out of bed with no lead dislodgment. CONCLUSION: The Tempo lead is safe and effective for temporary cardiac pacing for structural heart procedures, provides stable peri and postprocedural pacing and allows mobilization of patients who require temporary pacing leads.
No takes yet. Share an insight, caveat, or question.
Nazif et al. (2019) conducted an observational in Structural heart procedures requiring temporary cardiac pacing (n=269). Tempo active fixation temporary pacing lead was evaluated on Device safety (absence of clinically significant cardiac perforation, new pericardial effusion, or sustained ventricular arrhythmia) and efficacy (clinically acceptable pacing thresholds with successful pace capture). The Tempo temporary pacing lead was successfully positioned and achieved pacing in 98.1% of patients undergoing structural heart procedures, with no clinically significant perforations or effusions.