Answers are grounded in this paper and cite their sources.
Prospective diagnostic accuracy study evaluates IOP measurement reliability in children under 8 years, highlighting feasibility and correlation issues.
Objective To evaluate the feasibility, accuracy, and correlation of intraocular pressure (IOP) measurements obtained with the TD-8000 ® rebound tonometer compared with Perkins ® applanation tonometry in children with glaucoma or suspected glaucoma. Methods This prospective diagnostic accuracy study included 78 eyes of 44 children under 8 years. IOP was measured using the TD-8000 ® during outpatient visits and under general anesthesia. Perkins ® tonometry was performed under general anesthesia. Diagnostic performance was calculated using a 21 mmHg cutoff. Agreement was evaluated using correlation and Bland–Altman methods. Results Rebound tonometry was feasible in 89% of awake children. Under general anesthesia, TD-8000 ® overestimated IOP compared with Perkins ® , with a mean difference of 9.06 mmHg (SE = 0.62; p < 0.001). Accuracy was 87.5%, sensitivity 22.2%, specificity 98.2%, negative predictive value 88.5% and positive predictive value 66.7%. Outpatient rebound IOP measurements correlated weakly with applanation under general anesthesia (R = 0.32; p = 0.04). Conclusions The TD-8000 ® is a feasible but less accurate tool for IOP measurement in young children. Weak correlation between outpatient rebound tonometry and applanation under general anesthesia limits its reliability, although high specificity may offer clinical value in high-prevalence settings.
No takes yet. Share an insight, caveat, or question.
Belfort et al. (2026) studied this question.