Key result
A prototype fiber-optic immunobiosensing system quantified multiple anticoagulants within 5 minutes (S/N ratio 25) and four cardiac markers within 10 minutes (S/N ratio 20).
A prototype fiber-optic immunosensing system demonstrates rapid (<10 min) and simultaneous quantification of cardiovascular biomarkers, showing potential as a point-of-care diagnostic tool.
A prototype of a fiber-optic, multi-analyte, immunobiosensing system was developed to simultaneously quantify disease-representing biomarkers in blood plasma. This system was for simultaneous quantification of two different groups of multi-biomarkers related to cardiovascular diseases (CVD): anticoagulants (protein C, protein S, antithrombin III, and plasminogen) for deficiency diagnosis; and cardiac markers (B-type natriuretic peptide, cardiac troponin I, myoglobin, and C-reactive protein) for coronary heart disease diagnosis. As an initial effort towards the development of a disposable and easy-to-use sensing cartridge as a rapid diagnostic tool for CVD related diseases, a prototype of a flow control system was also developed to automatically perform simultaneous four-analyte quantification. Currently, the system is capable of quantifying the multiple anticoagulants in their clinically significant sensing ranges within 5 minutes, at an average signal-to-noise (S/N) ratio of 25. A simultaneous assay of the four cardiac markers can be performed within 10 min, at an average S/N ratio of 20. When this highly portable multi-analyte sensing system is completed and successfully tested for CVD patient's plasma, it can provide rapid (<10 min) and reliable diagnostic and prognostic information at a patient's bedside.
No takes yet. Share an insight, caveat, or question.
Tang et al. (2006) studied Cardiovascular diseases. Fiber-optic, multi-analyte, immunobiosensing system was evaluated on Simultaneous quantification of multiple anticoagulants and cardiac markers. A prototype fiber-optic immunobiosensing system quantified multiple anticoagulants within 5 minutes (S/N ratio 25) and four cardiac markers within 10 minutes (S/N ratio 20).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: