Key result
The Verigene CLO assay accurately identified homozygous and heterozygous CYP2C19 *2 and *3 phenotypes from whole blood with 100% specificity and a 99.7% final call rate in approximately 3.5 hours.
Why the study?
Does the Verigene CLO assay accurately identify CYP2C19 *2 and *3 polymorphisms in whole blood samples?
Cross-Sectional (n=1,286)
Does the Verigene CLO assay accurately identify CYP2C19 *2 and *3 polymorphisms in whole blood samples?
The Verigene CLO assay provides rapid and highly specific detection of CYP2C19 *2 and *3 polymorphisms from whole blood, which is relevant for guiding clopidogrel therapy.
Supports rapid point-of-care CYP2C19 genotyping for clopidogrel; leaves open outcome impact in prospective trials.
Numerous drugs such as clopidogrel have been developed to reduce coagulation or inhibit platelet function. The hepatic cytochrome P450 (CYP) pathway is involved in the conversion of clopidogrel to its active metabolite. A recent black-box warning was included in the clopidogrel package insert indicating a significant clinical link between specific CYP2C19 genetic variants and poor metabolism of clopidogrel. Of these variants, *2 and *3 are the most common and are associated with complete loss of enzyme activity. In patients who are carriers of a CYP2C19 *2 or *3 allele, the conversion of clopidogrel to its active metabolite may be reduced, which can lead to ischemic events and negative consequence for the patient. We examined the ability of the Verigene CLO assay (Nanosphere, Northbrook, IL) to identify CYP2C19 *2 and *3 polymorphisms in 1,286 unique whole blood samples. The Verigene CLO assay accurately identified homozygous and heterozygous *2 and *3 phenotypes with a specificity of 100% and a final call rate of 99.7%. The assay is fully automated and can produce a result in approximately 3.5 hours.
No takes yet. Share an insight, caveat, or question.
Buchan et al. (2011) conducted a cross-sectional in CYP2C19 genotyping for clopidogrel metabolism (n=1,286). Verigene CLO assay was evaluated on Identification of CYP2C19 *2 and *3 polymorphisms (specificity and final call rate). The Verigene CLO assay accurately identified homozygous and heterozygous CYP2C19 *2 and *3 phenotypes from whole blood with 100% specificity and a 99.7% final call rate in approximately 3.5 hours.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: