Answers are grounded in this paper and cite their sources.
It has been proposed that a thin diffuser can be used as a simple wavefront sensor. This study investigates the ability of such a sensor to estimate wavefronts corresponding to single or multiple low-order Zernike modes. We show that the accuracy of wavefront reconstruction depends on the specific Zernike mode within a wavefront, its amplitude, and the distance between the diffuser and detector. We find that the amplitude and type of aberrations within a wavefront determine the optimal detector–diffuser distance for wavefront reconstruction. Moreover, the distribution of individual amplitudes within a multi-mode containing wavefront significantly influences the accuracy of a thin diffuser wavefront sensor, with higher residual root-mean-square values observed when the magnitude of spherical aberration is greatest, followed by coma, astigmatism, and defocus, respectively.
No takes yet. Share an insight, caveat, or question.
Leporda et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: