Key result
The Biopac system measured consistently and significantly lower absolute pulse wave velocity values compared to the Complior system (3.0 vs 10.2 m/s in the sitting position, p<0.001), though the two methods showed good agreement in tracking changes.
Why the study?
Does the Biopac system provide comparable and correlating pulse wave velocity measurements to the Complior system in healthy volunteers?
Cross-Sectional (n=25)
No
Does the Biopac system provide comparable and correlating pulse wave velocity measurements to the Complior system in healthy volunteers?
While Biopac and Complior systems correlate well for measuring changes in pulse wave velocity, their absolute values differ significantly and cannot be used interchangeably for diagnosis.
Answers are grounded in this paper and cite their sources.
Biopac yields lower PWV than Complior yet tracks changes similarly; leaves open interchangeability for absolute-value clinical decisions.
Arterial stiffness is a reliable prognostic parameter for cardiovascular diseases. The effect of change in arterial stiffness can be measured by the change of the pulse wave velocity (PWV). The Complior system is widely used to measure PWV between the carotid and radial arteries by means of piezoelectric clips placed around the neck and the wrist. The Biopac system is an easier to use alternative that uses ECG and simple optical sensors to measure the PWV between the heart and the fingertips, and thus extends a bit more to the peripheral vasculature compared to the Complior system. The goal of this study was to test under various conditions to what extent these systems provide comparable and correlating values. 25 Healthy volunteers, 20-30 years old, were measured in four sequential position: sitting, lying, standing and sitting. The results showed that the Biopac system measured consistently and significantly lower PWV values than the Complior system, for all positions. Correlation values and Bland-Altman plots showed that despite the difference in PWV magnitudes obtained by the two systems the measurements did agree well. Which implies that as long as the differences in PWV magnitudes are taken into account, either system could be used to measure PWV changes over time. However, when basing diagnosis on absolute PWV values, one should be very much aware of how the PWV was measured and with what system.
No takes yet. Share an insight, caveat, or question.
Velzen et al. (2018) conducted a cross-sectional in Healthy volunteers (n=25). Biopac system vs. Complior system was evaluated on Pulse wave velocity (PWV) in the first sitting position (p=<0.001). The Biopac system measured consistently and significantly lower absolute pulse wave velocity values compared to the Complior system (3.0 vs 10.2 m/s in the sitting position, p<0.001), though the two methods showed good agreement in tracking changes.