Key result
Cuff-based PWV estimation shows greater short-term repeatability than carotid-femoral measurement.
Why the study?
The repeatability of aortic pulse wave velocity measures obtained with different non-invasive devices was not well characterized.
Cross-Sectional (n=102)
May caution against interchangeable PWV device use in practice; leaves open need for standardized validation protocols.
Objective: Aortic pulse wave velocity (PWV) is a validated indicator of central arterial stiffness and cardiovascular risk. We aimed to compare the repeatability of PWV measures obtained with non-invasive devices. Design and method: We evaluated the repeatibility of non-invasive measures of PWV, obtained with 4 devices measuring two-points carotid-femoral PWV (Complior, PulsePen ETT, PulsePen ET, SphygmoCor), and with 2 devices estimating PWV from the oscillometric cuff-derived brachial pulsewave (BPLab, Mobil-O-Graph). 102 patients planned to undertake a cardiac catheterization (age 65 ± 13 years, 70.6% males) were enrolled. Repeated measures of PWV were obtained with all devices in a single session, 15 minutes apart. Duplicate PWV and carotid-femoral PTT measurements were compared using different indices. Coefficients of variation (CV%) and their confidence intervals (CI) are reported. Results: Devices evaluating carotid-femoral PWV showed a good repeatability (CV%[CI] for Complior: 8.8[7.3–10.1]; PulsePen ETT: 8.0[6.2–9.5]; PulsePen ET: 5.8[4.9–6.6]; SphygmoCor: 9.5[7.7–11.0]), whereas the repeatability of PWV estimated by cuff-based devices was slightly higher (BPLab: 5.5[4.2–6.6], Mobil-O-Graph: 3.4[2.9–3.8]). A lower repeatability of carotid-femoral PWV was present for greater arterial stiffness values (CV%[CI] for PWV<10 m/s vs PWV>=10 m/s: Complior 7.0[5.4–8.3] vs 10.5[8.0–12.5], PulsePen ETT 6.3[3.6–8.1] vs 9.2[6.5–11.3], PulsePen ET 4.9[3.5–6.0] vs 6.5[5.3–7.6], Sphygmocor 8.5[5.7–10.6] vs 10.3[7.7–12.3]. No such difference was observed with cuff-based devices (BPLab 6.0[3.6–7.7] vs 5.1[3.5–6.4], Mobil-O-Graph 3.5[2.8–4.1] vs 3.2[2.6–3.7]). Differences between repeated PWV measurements were not correlated with concomitant blood pressure (R2: 0.005) or heart rate differences (R2: 0.013). Conclusions: Short-term repeatability of PWV measures was good but not homogenous among different devices. A greater repeatability was observed with cuff-based devices, compared to devices measuring carotid-femoral PWV. This is probably due for Mobil-O-Graph to the algorithm for PWV assessment, which considers age and mean blood pressure, and for BPLab to the automated editing procedure which eliminates highly variable PWV values. Repeatability of PWV is not influenced by blood pressure or heart rate concomitant changes. For carotid-femoral PWV, the repeatability of measures is lower for higher PWV values. These results could be usefully considered when assessing PWV in a clinical setting.
No takes yet. Share an insight, caveat, or question.
Grillo et al. (2017) conducted a cross-sectional in Patients planned to undertake a cardiac catheterization (n=102). Non-invasive PWV assessment devices (cuff-based vs carotid-femoral) vs. Repeated measures within and between devices was evaluated on Repeatability of pulse wave velocity (PWV) measures expressed as coefficient of variation (CV%). Cuff-based devices estimating pulse wave velocity showed greater short-term repeatability (CV% 3.4-5.5) compared to devices measuring carotid-femoral PWV (CV% 5.8-9.5).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: