Key result
The PP-1000 pulse wave velocity measurement system demonstrated high reproducibility in healthy subjects, with a between-observer correlation for aortic PWV of r=0.93 and within-observer correlations ranging from 0.94 to 0.99.
Why the study?
Does the PP-1000 pulse wave velocity measurement system provide reproducible regional PWV values in healthy male subjects?
Population
17 healthy male subjects without any cardiovascular disease, mean age 33 years (range 22-52).
Design
Cross-sectional, Two observers performed two consecutive measurements from the…
Authors
Loading...
Supports device validation in healthy subjects; leaves open clinical utility pending patient studies.
Observational (n=17)
No
Does the PP-1000 pulse wave velocity measurement system provide reproducible regional PWV values in healthy male subjects?
Effect estimate: r=0.93
p-value: p=<0.01
The PP-1000 system provides highly reproducible regional pulse wave velocity measurements in healthy subjects, supporting its potential for clinical application in evaluating arterial stiffness.
Lee et al. (2009) conducted an observational in Healthy (n=17). PP-1000 PWV measurement system vs. Between-observer and within-observer measurements was evaluated on Between-observer correlation coefficient for aortic PWV (r=0.93, p=<0.01). The PP-1000 pulse wave velocity measurement system demonstrated high reproducibility in healthy subjects, with a between-observer correlation for aortic PWV of r=0.93 and within-observer correlations ranging from 0.94 to 0.99.