Key result
The PP-1000 pulse wave velocity measurement system demonstrated high reproducibility in healthy subjects, with a between-observer correlation for aortic PWV of r=0.93 and within-observer correlations ranging from 0.94 to 0.99.
Why the study?
Does the PP-1000 pulse wave velocity measurement system provide reproducible regional PWV values in healthy male subjects?
Observational (n=17)
No
Does the PP-1000 pulse wave velocity measurement system provide reproducible regional PWV values in healthy male subjects?
The PP-1000 system provides highly reproducible regional pulse wave velocity measurements in healthy subjects, supporting its potential for clinical application in evaluating arterial stiffness.
Answers are grounded in this paper and cite their sources.
Supports device validation in healthy subjects; leaves open clinical utility pending patient studies.
BACKGROUND/AIMS: Despite the clinical importance and widespread use of pulse wave velocity (PWV), there are no standards for pulse sensors or for system requirements to ensure accurate pulse wave measurement. We assessed the reproducibility of PWV values using a newly developed PWV measurement system. METHODS: The system used in this study was the PP-1000, which simultaneously provides regional PWV values from arteries at four different sites (carotid, femoral, radial, and dorsalis pedis). Seventeen healthy male subjects without any cardiovascular disease participated in this study. Two observers performed two consecutive measurements in the same subject in random order. To evaluate the reproducibility of the system, two sets of analyses (within-observer and between-observer) were performed. RESULTS: The means+/-SD of PWV for the aorta, arm, and leg were 7.0+/-1.48, 8.43+/-1.14, and 8.09+/-0.98 m/s as measured by observer A and 6.76+/-1.00, 7.97+/-0.80, and 7.97+/-0.72 m/s by observer B, respectively. Betweenobserver differences for the aorta, arm, and leg were 0.14+/-0.62, 0.18+/-0.84, and 0.07+/-0.86 m/s, respectively, and the correlation coefficients were high, especially for aortic PWV (r=0.93). All the measurements showed significant correlation coefficients, ranging from 0.94 to 0.99. CONCLUSIONS: The PWV measurement system used in this study provides accurate analysis results with high reproducibility. It is necessary to provide an accurate algorithm for the detection of additional features such as flow wave, reflection wave, and dicrotic notch from a pulse waveform.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2009) conducted an observational in Healthy (n=17). PP-1000 PWV measurement system vs. Between-observer and within-observer measurements was evaluated on Between-observer correlation coefficient for aortic PWV (r=0.93, p=<0.01). The PP-1000 pulse wave velocity measurement system demonstrated high reproducibility in healthy subjects, with a between-observer correlation for aortic PWV of r=0.93 and within-observer correlations ranging from 0.94 to 0.99.