Key result
Burst antitachycardia pacing therapy showed no significant difference in success compared to ramp therapy for spontaneous ventricular tachycardia (OR 1.116; 95% CI 0.788-1.579).
Why the study?
Current guidelines prefer burst over ramp antitachycardia pacing therapy for ventricular tachycardia, but the comparative success and acceleration rates remained to be determined.
Does burst ATP therapy improve success or reduce acceleration compared to ramp ATP therapy in patients with ventricular tachycardia?
Meta-Analysis (n=1,672)
Does burst ATP therapy improve success or reduce acceleration compared to ramp ATP therapy in patients with ventricular tachycardia?
Burst and ramp antitachycardia pacing therapies demonstrate similar success and acceleration rates for treating ventricular tachycardia, suggesting neither should be preferred over the other.
No difference in ATP success supports flexible ICD programming; challenges burst preference and leaves optimal strategy open.
Current guidelines recommend at least one attempt of defibrillator antitachycardia pacing (ATP) therapy, showing preference for burst therapy. The objective of this study is to compare ramp versus burst ATP therapy proportion of success and acceleration in treating spontaneous or induced ventricular tachycardia (VT). The review protocol was previously published in PROSPERO. Data synthesis and measures of heterogeneity ( I 2 ) was performed by CMA® software v.3 comparing proportions in both groups. Sensitivity analysis was performed as subgroup or meta‐regression according to quality, clinical characteristics, and differences in design. Thirteen studies including 30,117 VT episodes in 1672 patients were analyzed. There was no significant difference in the proportion of success between burst and ramp therapy in spontaneous VT (odds ratio = 1.116; 95% confidence interval [CI] = 0.788–1.579; I 2 = 89%). There was no significant difference in the proportion of success between burst and ramp therapy in induced VT (odds ratio = 0.820; 95% CI = 0.468–1.437; I 2 = 93%). No significant difference was found in the proportion of acceleration between burst and ramp in spontaneous VT (odds ratio = 0.792; 95% CI = 0.476–1.317; I 2 = 83%). No significant difference was found in the proportion of acceleration between burst and ramp in induced VT (odds ratio = 1.234; 95% CI = 0.802–1.898; I 2 = 55%). Sensitivity analysis did not change main results. There is no difference in success or in acceleration proportion between burst or ramp ATP therapy irrespective if the VT was spontaneous or induced. Future implantable cardioverter defibrillator programming guidelines should offer both ATP therapies without preference in one of them.
No takes yet. Share an insight, caveat, or question.
Sousa et al. (2021) conducted a meta-analysis in Ventricular tachycardia (n=1,672). Burst antitachycardia pacing (ATP) therapy vs. Ramp antitachycardia pacing (ATP) therapy was evaluated on Proportion of success in treating spontaneous ventricular tachycardia (OR 1.116, 95% CI 0.788-1.579). Burst antitachycardia pacing therapy showed no significant difference in success compared to ramp therapy for spontaneous ventricular tachycardia (OR 1.116; 95% CI 0.788-1.579).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: