Key result
Doppler velocity measurements accurately quantified aortic flow (r = 0.98-0.99), and variations in sample length and position did not produce significant differences in calculated flows.
Why the study?
Do variations in pulsed Doppler sampling site affect the calculation of cardiac output in an open-chest dog model?
Do variations in pulsed Doppler sampling site affect the calculation of cardiac output in an open-chest dog model?
Effect estimate: r = 0.98-0.99
Variations in Doppler sample length and position do not significantly affect the accuracy of calculated aortic flows in a laminar flow model.
Sample volume variations may affect aortic flow accuracy; leaves open standardized protocols for clinical echocardiography.
We measured aortic flow by two-dimensional Doppler echocardiography in an open-chest dog model to examine how variations in Doppler sample volume length and position influence aortic hemodynamic flow calculations. Fourteen dogs underwent right-heart bypass, in which venous return from the venae cavae drained by gravity to a reservoir. A variable-speed roller pump returned the blood to the pulmonary artery, fixing left-sided cardiac input and output. Echo Doppler measurements were performed using a 3.5 MHz transducer placed directly on the aortic arch to determine internal aortic cross-sectional area. The transducer was then directed to image the aortic arch for Doppler velocity measurements and the various sampling sites were investigated. Doppler cardiac output could then be determined for each of the various sample volumes over a range of known roller pump settings. Doppler velocity was analyzed using fast Fourier transform spectral analysis. Mean velocity over the cardiac cycle was obtained by planimetry of the area under the Doppler velocity curve with a minicomputer. Doppler-derived determinations of cardiac output achieved a correlation of r = 0.98-0.99 to values obtained by the roller pump over a range of cardiac outputs from 0.75-5 l/min. The standard error of the estimate was 0.21/min. In this laminar flow model, there was no difference between the predictive accuracy of any of the sampling sites over the range of roller pump flows. Our study shows that Doppler velocity measurements can be used to quantify aortic flow over a clinically useful range and that variations of sample length and position did not produce significant differences in calculated flows.
No takes yet. Share an insight, caveat, or question.
Fisher et al. (1983) studied this question. Variations in Doppler sample volume length and position vs. Known roller pump settings was evaluated on Correlation of Doppler-derived cardiac output to values obtained by the roller pump (r = 0.98-0.99). Doppler velocity measurements accurately quantified aortic flow (r = 0.98-0.99), and variations in sample length and position did not produce significant differences in calculated flows.
Synapse has enriched one closely related paper. Consider it for comparative context: