Answers are grounded in this paper and cite their sources.
A novel procedure for making the pH responsiveness of a Gd-based probe independent of its concentration has been developed. The method is based on a ratiometric approach and requires the measurement, at magnetic fields higher than 1 T, of the paramagnetic contribution to the longitudinal and the transverse NMR relaxation rates of water protons in the presence of a Gd-based probe with a rotational mobility in the nanosecond time scale. The method has been tested and validated in vitro by using a macromolecular adduct based on the polyornithine backbone, whose rotational mobility is pH dependent. The results reported in this communication improve substantially the diagnostic potential of MRI responsive agents.
No takes yet. Share an insight, caveat, or question.
Aime et al. (2006) studied this question.