Key result
Real-time MR-guided radiofrequency ablation in an isolated beating pig heart produced focal T1 reductions (936 ± 80 ms) vs nonablated myocardium (1192 ± 26 ms), consistent with necrosis.
Why the study?
Interventional cardiovascular MRI guidance for myocardial ablation faces development limits due to challenges in MR-compatible instrument testing, sequence validation, and histopathologic correlation hindered by in vivo tissue evaluation limitations.
Does real-time MR-guided radiofrequency ablation produce characterizable lesions in an isolated beating pig heart platform?
Does real-time MR-guided radiofrequency ablation produce characterizable lesions in an isolated beating pig heart platform?
Real-time MR-guided radiofrequency ablation is feasible in an isolated beating pig heart platform, enabling high-resolution lesion assessment and histologic correlation for future MR-guided electrophysiology therapies.
Feasibility of real-time MR-guided ablation shown in isolated beating heart model; leaves open translation to clinical use pending prospective validation.
Background: Interventional cardiovascular magnetic resonance imaging (MRI) offers real-time, radiation-free guidance for complex procedures such as myocardial ablation, marking a promising advance in electrophysiology. However, further development is limited by challenges in magnetic resonance (MR)-compatible instrument testing, MRI sequence validation, and accurate correlation with histopathology, hindered by the limitations of in vivo tissue evaluation. Objective: This study investigated the feasibility of real-time MR-guided radiofrequency (RF) ablation in an MR-compatible isolated beating pig heart platform and characterized ablation lesions using MRI and histopathology. Methods: A heart from a pig slaughtered for human consumption was prepared under regulatory guidelines and connected to a custom-built, MR-compatible perfusion platform supporting left ventricular function in both Langendorff and working modes. Autologous heparinized blood circulated at physiological pressures and temperatures. MR-guided catheter navigation and RF ablation were performed on a Philips 3T scanner using active catheter tracking. Native T1 and T2 mapping were acquired before and after ablation. Lesions were confirmed by histologic analysis. Results: RF ablation (50 W, 60 seconds) was successfully performed at 5 left ventricular sites. MRI showed focal reductions in T1 (936 ± 80 ms) surrounded by elevated T1 (1357 ± 18 ms) and T2 values (86 ± 10 ms) compared with nonablated myocardium (T1 1192 ± 26 ms; T2 66 ± 6 ms), consistent with necrosis and edema. Histology confirmed a necrotic core with a surrounding rim showing contraction band necrosis and erythrocyte extravasation. Conclusion: This study demonstrates the feasibility of real-time MR-guided ablation in a beating pig heart platform. The setup allows high-resolution lesion assessment and histologic correlation, supporting future developments in MR-guided therapies.
No takes yet. Share an insight, caveat, or question.
Hopman et al. (2025) studied this question. Real-time MR-guided radiofrequency ablation vs. Nonablated myocardium was evaluated on Lesion characterization using MRI (T1 and T2 mapping) and histopathology. Real-time MR-guided radiofrequency ablation in an isolated beating pig heart produced focal T1 reductions (936 ± 80 ms) vs nonablated myocardium (1192 ± 26 ms), consistent with necrosis.