An experimental arrangement is described which was set up to measure the energy loss spectra of electrons which have passed through thin specimens in the electron microscope. It consists of a nondispersive filter lens which rejects electrons of less than a certain energy, determined by the voltage of the central electrode. By modulating this voltage with a small ac component and measuring the resulting modulation of the transmitted beam, the apparatus can be used as an energy analyser. In this paper, the process of `differentiation by modulation' is reviewed and the effects of using different waveforms and frequencies examined. Also discussed are high-voltage insulation and breakdown in vacuum, the effect of focusing the beam on the phosphor (which was used as the electron detector), and limitations upon possible sweep speeds. A novel feature of the system is a 2 kHz optical link used to supply a synchronizing signal to the high-voltage line.