Answers are grounded in this paper and cite their sources.
Photoacoustic imaging is an emerging noninvasive and nonionizing imaging modality with high contrast and high resolution, which originates from the photoabsorption selectivity and the broadband response of the photoacoustic signals. However, sensing of the photoacoustic signal is currently limited by the bandwidth of piezoelectric ultrasound transducers. The optical ultrasound sensors based on Fabry–Perot (F–P) interference exhibit broader bandwidth than piezoelectric transducers, which is expected to be a promising alternative. At present, nevertheless, the parallel sensing and processing of ultrasound signals using the fiber-optic ultrasound sensor is still a challenge. Here, we propose a <tex-math notation="LaTeX">4×16 </tex-math> fiber-optic array based on F–P interference, which enables parallel sensing for imaging with a volume rate of 10 Hz. We experimentally demonstrated that the fiber-optic array ultrasound sensor exhibited a pressure nonlinearity of <3.64%, and −6 dB bandwidth from ~0 to 18.3 MHz. High-definition imaging is achieved by this sensing array without mechanical scanning. As the ultrasound sensing realized by optics, the photoacoustic imaging system is potentially an all-optic system with broadband response.
No takes yet. Share an insight, caveat, or question.
Ma et al. (2022) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: