Key result
During induced ventricular fibrillation in S-ICD patients, mean time to first therapy was 16.2 seconds, with only 29% showing optimal detection and 6% having therapy inhibited by noise oversensing.
Why the study?
Does intraoperative defibrillation testing identify sensing delays or failures in patients receiving an S-ICD?
Observational (n=137)
Yes
Does intraoperative defibrillation testing identify sensing delays or failures in patients receiving an S-ICD?
Answers are grounded in this paper and cite their sources.
Intraoperative defibrillation testing during S-ICD implantation frequently reveals marked sensing delays and occasional noise oversensing that inhibits therapy, supporting the need for systematic testing.
BackgroundThe subcutaneous implantable cardioverter–defibrillator (S-ICD) remains a new technology requiring accurate assessment of the various aspects of its functioning. Isolated cases of delayed sensing of ventricular arrhythmia have been described.ObjectiveThe purpose of this multicenter study was to assess the quality of sensing during induced ventricular fibrillation (VF).MethodsOne hundred thirty-seven patients underwent induction of VF at the end of the S-ICD implantation.ResultsVF induction was successful in 133 patients (97%). Mean time to first therapy was 16.2 ± 3.1 seconds, with a substantial range from 12.5 to 27.0 seconds. Four different detection profiles were arbitrarily defined: (1) optimal detection (n = 39 [29%]); (2) undersensing with moderate prolongation of time to therapy (<18 seconds; n = 68 [51%]); (3) undersensing with significant prolongation of the time to therapy (>18 seconds; n = 19 [14%]); and (4) absence of therapy or prolonged time to therapy related to noise oversensing (n = 7 [6%]). In some of the patients in the last group, despite induction of VF the initial counter was never filled, the device did not charge the capacitors, and the shock was not delivered because of a sustained diagnosis of noise (n = 5). A manual shock by the device or an external shock had to be delivered to restore the sinus rhythm.ConclusionOur study demonstrated a marked sensing delay leading to prolonged time to therapy in a large number of S-ICD patients. A few worrisome cases of noise oversensing inhibiting the therapies were detected. These results support the need for systematic intraoperative defibrillation testing. The subcutaneous implantable cardioverter–defibrillator (S-ICD) remains a new technology requiring accurate assessment of the various aspects of its functioning. Isolated cases of delayed sensing of ventricular arrhythmia have been described. The purpose of this multicenter study was to assess the quality of sensing during induced ventricular fibrillation (VF). One hundred thirty-seven patients underwent induction of VF at the end of the S-ICD implantation. VF induction was successful in 133 patients (97%). Mean time to first therapy was 16.2 ± 3.1 seconds, with a substantial range from 12.5 to 27.0 seconds. Four different detection profiles were arbitrarily defined: (1) optimal detection (n = 39 [29%]); (2) undersensing with moderate prolongation of time to therapy (<18 seconds; n = 68 [51%]); (3) undersensing with significant prolongation of the time to therapy (>18 seconds; n = 19 [14%]); and (4) absence of therapy or prolonged time to therapy related to noise oversensing (n = 7 [6%]). In some of the patients in the last group, despite induction of VF the initial counter was never filled, the device did not charge the capacitors, and the shock was not delivered because of a sustained diagnosis of noise (n = 5). A manual shock by the device or an external shock had to be delivered to restore the sinus rhythm. Our study demonstrated a marked sensing delay leading to prolonged time to therapy in a large number of S-ICD patients. A few worrisome cases of noise oversensing inhibiting the therapies were detected. These results support the need for systematic intraoperative defibrillation testing.
No takes yet. Share an insight, caveat, or question.
Waroux et al. (2018) conducted an observational in Subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation (n=137). Subcutaneous implantable cardioverter-defibrillator (S-ICD) was evaluated on Quality of sensing and time to first therapy during induced ventricular fibrillation. During induced ventricular fibrillation in S-ICD patients, mean time to first therapy was 16.2 seconds, with only 29% showing optimal detection and 6% having therapy inhibited by noise oversensing.