Key result
A validated LC-MS/MS method for determining rivaroxaban in human plasma demonstrated high accuracy (96.3-102%) and precision (1.08-3.75%) and was successfully applied to a pharmacokinetic study.
A validated LC-MS/MS method for quantifying rivaroxaban in human plasma demonstrated high precision and accuracy, making it suitable for pharmacokinetic studies.
Validated LC-MS/MS method quantifies rivaroxaban in plasma; extends bioanalytical options but leaves open clinical adoption.
A high-performance liquid chromatography tandem mass spectrometric method for the determination of Rivaroxaban in human plasma has been developed and validated using Rivaroxaban D 4 as an internal standard. The extraction of analyte and internal standard was accomplished by solid phase extraction technique. The method has been validated over a concentration range of 5.96–801 ng/mL. Chromatographic separations were achieved using Gemini C18, 150 mm × 4.6 mm, 5 µm, column eluted at flow rate of 1.5 mL/min with mobile phase (acetonitrile: ammonium acetate buffer (80:20 v/v)). The overall run time of method was about 1.8 min with elution times of Rivaroxaban and its internal standard Rivaroxaban D 4 at around 1.18 min. The multiple reaction monitoring transitions were set at 436/145 (m/z) and 440/145 (m/z) for Rivaroxaban and Rivaroxaban D 4 , respectively. The calibration curves were linear (r 2 ≥ 0.99) over the range of 5.96–801 ng/mL with lower limit of quantitation validated at 5.96 ng/mL. Extraction recoveries were >88% for both rivaroxaban and its stable labeled internal standard rivaroxaban D 4 . The inter-day/between run precisions were ranged from 1.08% to 3.75%, while accuracy ranged from 96.3% to 102%. The presented method was used in pharmacokinetic study in healthy volunteers. Results of incurred sample reanalysis were within the acceptance range of ±20% of original value, for 98.3% of samples reanalyzed. This indicated good assay precision of target analytes in their real matrix at the employed experimental conditions. The applicability of the assay for the determination of the pharmacokinetic parameters was demonstrated.
No takes yet. Share an insight, caveat, or question.
Shaikh et al. (2019) studied Healthy volunteers. Rivaroxaban was evaluated on Method validation (accuracy, precision, recovery) and pharmacokinetic parameters. A validated LC-MS/MS method for determining rivaroxaban in human plasma demonstrated high accuracy (96.3-102%) and precision (1.08-3.75%) and was successfully applied to a pharmacokinetic study.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: