Answers are grounded in this paper and cite their sources.
In this work, a validated stability-indicating ultra-high performance liquid chromatographic (UPLC) method has been developed for quantitative determination of related substances in Risperdalâ Constaâ. The chromatographic separation was achieved on a Waters BEH C18 column (2.1 × 100 mm, 1.7 μm) in an isocratic elution mode. The limit of quantification is 0.1 μg/mL, and the method is linear over the concentration range of 0.1-1.5 μg/mL for risperidone. The PLGA polymers have been dissolved in organic phase (acetonitrile) and precipitated in aqueous phase (0.01 N HCl). This simple step avoids the clogging of column and makes the method more robust. The mean recovery of extracted risperidone from PLGA microspheres is 99.26 ± 1.22%. Herein, this method was proved to be highly reliable and applicable for measuring impurities in PLGA microspheres.
No takes yet. Share an insight, caveat, or question.
Cheng et al. (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: