Key result
Five wearable ECG devices demonstrated high QRS detection performance across rest and movement conditions, with positive predictive values between 0.985 and 1.000.
Why the study?
Numerous wearable devices are used to record cardiac activity, but their validity and usability across different realistic use cases have not been fully investigated.
How do five different wearable ECG devices compare in signal quality and QRS detection accuracy during various physical and cognitive tasks?
Observational (n=13)
How do five different wearable ECG devices compare in signal quality and QRS detection accuracy during various physical and cognitive tasks?
Five wearable ECG devices demonstrated high QRS detection accuracy during various tasks, with the eMotion Faros 360° providing the most stable signal quality.
Wearable ECGs may enable reliable monitoring in low-motion settings; hypothesis-generating for device selection and broader validation.
BACKGROUND: Numerous wearables are used in a research context to record cardiac activity although their validity and usability has not been fully investigated. The objectives of this study is the cross-model comparison of data quality at different realistic use cases (cognitive and physical tasks). The recording quality is expressed by the ability to accurately detect the QRS complex, the amount of noise in the data, and the quality of RR intervals. METHODS: Five ECG devices (eMotion Faros 360°, Hexoskin Hx1, NeXus-10 MKII, Polar RS800 Multi and SOMNOtouch NIBP) were attached and simultaneously tested in 13 participants. Used test conditions included: measurements during rest, treadmill walking/running, and a cognitive 2-back task. Signal quality was assessed by a new local morphological quality parameter morphSQ which is defined as a weighted peak noise-to-signal ratio on percentage scale. The QRS detection performance was evaluated with eplimited on synchronized data by comparison to ground truth annotations. A modification of the Smith-Waterman algorithm has been used to assess the RR interval quality and to classify incorrect beat annotations. Evaluation metrics includes the positive predictive value, false negative rates, and F1 scores for beat detection performance. RESULTS: All used devices achieved sufficient signal quality in non-movement conditions. Over all experimental phases, insufficient quality expressed by morphSQ values below 10% was only found in 1.22% of the recorded beats using eMotion Faros 360°whereas the rate was 8.67% with Hexoskin Hx1. Nevertheless, QRS detection performed well across all used devices with positive predictive values between 0.985 and 1.000. False negative rates are ranging between 0.003 and 0.017. eMotion Faros 360°achieved the most stable results among the tested devices with only 5 false positive and 19 misplaced beats across all recordings identified by the Smith-Waterman approach. CONCLUSION: Data quality was assessed by two new approaches: analyzing the noise-to-signal ratio using morphSQ, and RR interval quality using Smith-Waterman. Both methods deliver comparable results. However the Smith-Waterman approach allows the direct comparison of RR intervals without the need for signal synchronization whereas morphSQ can be computed locally.
No takes yet. Share an insight, caveat, or question.
Bläsing et al. (2022) reported an observational. Five wearable ECG devices was evaluated on QRS detection performance and signal quality. Five wearable ECG devices demonstrated high QRS detection performance across rest and movement conditions, with positive predictive values between 0.985 and 1.000.