Here, we developed a novel high-performance liquid chromatography (HPLC) method for quantification of perampanel in clinical practice and investigated the relationships between the plasma concentrations of perampanel obtained by this HPLC method and the CYP3A4*1G polymorphism. The developed HPLC method was validated based on US Food and Drug Administration. The developed HPLC method could be performed with a plasma volume of only 200 μL and had a limit of quantification (LOQ) of 2.5 ng/mL. The coefficients of variation (CVs) for intra- and inter-day assays were less than 10.4 and 7.2%, respectively, and the accuracy was <2.4% for both assays. A total of 12 patients who received 2 mg perampanel had C0 values ranging from 70.5 to 451 ng/mL, and the CV showed a large variation of 51.4%. No correlations were observed between the dose-adjusted C0 and the CYP3A4*1G polymorphism. This method was superior to previously reported methods in terms of plasma volume and LOQ and was clinically applicable. Perampanel showed high variations in individual plasma concentrations; however, individual differences could not be predicted from analysis of the CYP3A4*1G polymorphism before perampanel administration. Therefore, after beginning perampanel treatment, the dose should be determined based on the observed plasma concentration.