Key result
A 0/1 h diagnostic algorithm using a high-sensitivity troponin I point-of-care assay achieved a negative predictive value of 99.7% for ruling out non-ST-elevation myocardial infarction.
Why the study?
Increasing emergency department visits for suspected myocardial infarction require faster decision-making to reduce facility burden, prompting evaluation of a point-of-care high-sensitivity troponin I assay.
Does a 0/1 h diagnostic algorithm using a high-sensitivity troponin I point-of-care assay accurately diagnose non-ST-elevation myocardial infarction in patients presenting to the emergency department with suspected myocardial infarction?
Cohort (n=1,279)
Does a 0/1 h diagnostic algorithm using a high-sensitivity troponin I point-of-care assay accurately diagnose non-ST-elevation myocardial infarction in patients presenting to the emergency department with suspected myocardial infarction?
A point-of-care high-sensitivity troponin I assay using a 0/1 hour algorithm safely and effectively rules out or rules in NSTEMI with performance comparable to laboratory-based assays.
Answers are grounded in this paper and cite their sources.
Supports rapid NSTEMI rule-out with hs-TnI POC assay; leaves open generalizability pending prospective validation.
BACKGROUND: Increasing numbers of patients are presenting worldwide to emergency departments with suspected myocardial infarction. The use of point-of-care troponin assays might enable faster decision-making in this high-risk population and reduce the burden on emergency facilities. Here, we evaluate the diagnostic performance of a point-of-care high-sensitivity troponin I assay. METHODS: We conducted a prospective cohort study including patients presenting to the emergency department with suspected myocardial infarction from July 2013 to July 2016. A diagnostic algorithm for a high-sensitivity troponin I point-of-care assay was developed in a derivation data set with 669 patients and validated in an additional 610 patients. RESULTS: The derived 0/1 h algorithm for the point-of-care assay consisted of an admission troponin I <4 ng/L and a δ from 0 h to 1 h <3 ng/L for rule out and an admission troponin I ≥90 ng/L or a δ from 0 h to 1 h ≥20 ng/L for rule in of non-ST-elevation myocardial infarction. Application to the validation cohort showed a negative predictive value of 99.7% (95% CI, 98.1%-100.0%) and 48.0% of patients ruled out, whereas 14.6% were ruled in with a positive predictive value of 86.5% (95% CI, 77.6%-92.8%). The diagnostic performance of the point-of-care high-sensitivity assay was highly comparable to guideline-recommended use of a laboratory-based high-sensitivity troponin assay. CONCLUSIONS: The clinical application of a 0/1 h diagnostic algorithm based on a high-sensitivity troponin I point-of-care assay is safe, and diagnostic performance is comparable to a laboratory-based high-sensitivity troponin I assay.
No takes yet. Share an insight, caveat, or question.
Sörensen et al. (2019) conducted a cohort in Suspected myocardial infarction (n=1,279). High-sensitivity troponin I point-of-care assay 0/1 h algorithm vs. Laboratory-based high-sensitivity troponin assay was evaluated on Negative predictive value for rule out of non-ST-elevation myocardial infarction (95% CI 98.1-100.0). A 0/1 h diagnostic algorithm using a high-sensitivity troponin I point-of-care assay achieved a negative predictive value of 99.7% for ruling out non-ST-elevation myocardial infarction.