Key result
VEC-MRI detects a ~17-percentage-point increase in in-stent pulmonary regurgitant fraction over 3 months.
Answers are grounded in this paper and cite their sources.
BACKGROUND: Patients with endovascular stent implantation for the treatment of right ventricular outflow tract obstruction are often left with incomplete relief of the obstruction and significant pulmonary regurgitation. A noninvasive and reproducible method for monitoring such patients is desirable. MRI in the presence of a stent, however, has to overcome the problem of potential metallic artifacts. METHODS AND RESULTS: Under x-ray fluoroscopic guidance, endovascular nitinol stents were placed across the pulmonary valve in 6 young pigs to induce pulmonary regurgitation. Five additional pigs served as controls. Initial MRI was performed after 2 days (13.5+/-1.8 kg) and follow-up after 3 months (32+/-2.9 kg). Pulmonary flow volumes and regurgitant fraction were quantified by velocity-encoded cine (VEC) MRI through (VEC-TS) and distal to (VEC-DS) the stent. VEC-TS was compared with VEC-DS and volumetric measurements of left and right ventricular stroke volumes provided by cine MRI ("gold standard"). Antegrade and retrograde pulmonary flow volumes by VEC-TS were slightly but significantly less than those with VEC-DS and cine MRI. Excellent correlations (r>0.97) for phasic pulmonary flow volumes as measured by VEC-TS and VEC-DS were shown. Pulmonary regurgitant fraction increased from 32.8+/-15% to 49.6+/-17% (P<0.05) over the course of 3 months with VEC-TS. CONCLUSIONS: MRI demonstrates the progression of pulmonary regurgitation in growing swine. VEC MRI has the ability to quantify pulmonary blood flow inside the lumen of nitinol stents. MRI appears to be ideally suited for monitoring patients with endovascular nitinol stents in the pulmonary artery or pulmonary valve position.
No takes yet. Share an insight, caveat, or question.
Kühne et al. (2001) studied Pulmonary regurgitation (n=11). Velocity-encoded cine (VEC) MRI vs. Distal VEC (VEC-DS) and cine MRI (gold standard) was evaluated on Change in pulmonary regurgitant fraction over 3 months as measured by VEC-TS (p=<0.05). Velocity-encoded cine MRI successfully quantified pulmonary flow inside nitinol stents, demonstrating an increase in pulmonary regurgitant fraction from 32.8% to 49.6% over 3 months in growing swine.
Synapse has enriched 10 related papers. Consider them for comparative context: