Key result
Self-expanding prostheses had a higher eccentricity index (15.0% vs 7.1%, p<0.001) and lower expansion (68.3% vs 82.8%, p<0.001) compared to balloon-expandable prostheses at 3 months post-TAVR.
Why the study?
Does self-expanding prosthesis compared to balloon-expandable prosthesis affect prosthesis geometry and paravalvular regurgitation in patients undergoing TAVR?
Observational (n=147)
Does self-expanding prosthesis compared to balloon-expandable prosthesis affect prosthesis geometry and paravalvular regurgitation in patients undergoing TAVR?
Absolute Event Rate: 15% vs 7.1%
p-value: p=< .001
Self-expanding TAVR prostheses exhibit greater eccentricity and under-expansion compared to balloon-expandable valves, factors that are associated with the development of paravalvular regurgitation.
Differences in post-TAVR prosthesis geometry may inform device choice; leaves open causal link to PVR in randomized settings.
OBJECTIVES: Compare final morphology of self-expanding and balloon-expandable prosthesis and association with paravalvular regurgitation (PVR). BACKGROUND: PVR after transcatheter aortic valve replacement (TAVR) remains a frequent complication. A better understanding of the prosthesis geometry may be important to improve selection of the best device for each case and possibly reduce the rates of PVR. METHODS: Retrospective study including patients consecutively submitted to transcatheter aortic valve replacement: August/2007-October/2016. Three months after the procedure a multidetector computed tomography (MDCT) was performed to assess prosthesis geometry: dimensions, eccentricity, and expansion. RESULTS: A total of 147 individuals were included (mean age of 78.8 ± 6.7 and 50.3% males), 57% treated with a self-expanding prosthesis. On the postprocedure MDCT, the self-expanding group had higher eccentricity index (15.0 vs. 7.1%, p < .001) and lower expansion (68.3 vs. 82.8%, p < .001). In that group, the volume of calcium of landing zone had a significant correlation with eccentricity index and under-expansion. Patients with ≥mild PVR presented higher eccentricity (12.6 vs. 7.9%, p < .001) and lower expansion (68 vs. 75%, p = .012). Eccentricity index and landing zone calcium volume were independent predictors of PVR. CONCLUSIONS: Self-expanding prosthesis have greater eccentricity and under-expansion. Calcium burden exerts more influence in the final morphology of that type of valve. Calcification and eccentricity are associated with the development of PVR. These factors should be considered in the selection of the most appropriate type of prosthesis for each scenario.
No takes yet. Share an insight, caveat, or question.
Almeida et al. (2017) conducted an observational in Transcatheter aortic valve replacement (TAVR) (n=147). Self-expanding prosthesis vs. Balloon-expandable prosthesis was evaluated on Eccentricity index on postprocedure MDCT (p=< .001). Self-expanding prostheses had a higher eccentricity index (15.0% vs 7.1%, p<0.001) and lower expansion (68.3% vs 82.8%, p<0.001) compared to balloon-expandable prostheses at 3 months post-TAVR.