Key result
Ultrasound measurements of carotid stenosis showed considerable intersonographer and intermethod variability, with limits of agreement varying between +/-19.7 and 26.5%.
Why the study?
What is the intersonographer reproducibility and intermethod variability of ultrasound measurements of carotid artery stenosis in patients with stenotic carotid arteries?
Cross-Sectional (n=51)
What is the intersonographer reproducibility and intermethod variability of ultrasound measurements of carotid artery stenosis in patients with stenotic carotid arteries?
Considerable intersonographer and intermethod variability exists in ultrasound measurements of carotid stenosis, particularly in low-grade stenosis, which could impact clinical decision-making.
Answers are grounded in this paper and cite their sources.
Caution is warranted when relying on carotid ultrasound stenosis measurements for decisions; leaves open the need for standardized protocols in future studies.
BACKGROUND AND PURPOSE: Knowledge of the reproducibility of a diagnostic method is important in order to evaluate its usefulness. Few studies have examined interobserver and intermethod agreement on ultrasound measurements of carotid stenosis. METHODS: Intersonographer agreement on ultrasound measurements of carotid plaque morphology and the estimated degree of stenosis by three ultrasound methods were assessed in a random sample of 51 participants with stenotic carotid arteries selected from a population health survey. The degree of stenosis was assessed by measurements of velocity, lumen diameter reduction and cross-sectional lumen area. Intermethod agreement on the degree of carotid stenosis was also assessed. RESULTS: Agreement on plaque echogenicity and heterogeneity was moderate (kappa = 0.56 and kappa = 0.60, respectively). The mean degree of stenosis and median absolute difference between observers of the estimated degree of stenosis by the velocity method were 46.3 and 10.8%, respectively. The corresponding values were 51.0 and 5.8% for the diameter method, and 57.1 and 7.2%, for the cross-sectional lumen method. The limits of agreement for intersonographer reproducibility varied between +/-19.7 and 26.5%. For all methods, reproducibility increased with increasing degree of stenosis. Differences between the methods were large in low-grade stenosis but were acceptable in high-grade stenosis. CONCLUSIONS: Considerable differences in ultrasound measurement of stenosis, which could lead to different clinical conclusions, were regularly encountered no matter what ultrasound method was used.
No takes yet. Share an insight, caveat, or question.
Mathiesen et al. (2000) conducted a cross-sectional in Stenotic carotid arteries (n=51). Ultrasound measurements of carotid stenosis vs. Different sonographers and ultrasound methods (velocity, diameter, cross-sectional area) was evaluated on Intersonographer agreement on ultrasound measurements of carotid plaque morphology and estimated degree of stenosis. Ultrasound measurements of carotid stenosis showed considerable intersonographer and intermethod variability, with limits of agreement varying between +/-19.7 and 26.5%.
Synapse has enriched 10 related papers. Consider them for comparative context: