Key result
The FD-3 digital Holter recorder showed positive correlation with the Predictor system for signal-averaged ECG, with 91% concordance in late potential classification at 40-250 Hz.
Why the study?
Does a new digital Holter recording system (FD-3) provide comparable signal-averaged ECG analysis to an established bedside real-time system in patients undergoing ECG monitoring?
Observational (n=22)
Does a new digital Holter recording system (FD-3) provide comparable signal-averaged ECG analysis to an established bedside real-time system in patients undergoing ECG monitoring?
The FD-3 digital Holter recorder is feasible for signal-averaged ECG analysis and shows high concordance with established bedside systems, particularly at 40-250 Hz.
Answers are grounded in this paper and cite their sources.
May enable cost-effective Holter analysis; leaves open validation against established systems in larger cohorts.
Background: The implementation of signal‐averaged analysis in the Holter recording may increase cost‐effectiveness and enables further insights into pathophysiological links between arrhythmia genesis and time related modulating factors. This study assesses the feasibility of a new digitizing long‐term recorder system to perform a signal‐averaged analysis and compares the data with the results achieved by an established bedside real‐time recording system. Methods: The study was performed prospectively in 22 patients. The digital recorder FD‐3 obtained a 3‐channel 24‐hour Holter monitoring. For the signal‐averaged ECG analysis, a template averaging was achieved in 5‐minute periods. The interpolated sampling rate was 1024 Hz. The signal‐averaged ECG by bedside real‐time system was performed using the Predictor system. At 25–250 Hz and 40–250 Hz band‐pass filter, 4 FD‐3 epochs acquired in a 6‐hour interval were chosen and the mean values of QRS, root mean square (RMS), and low amplitude signal (LAS) were compared with the data obtained by one Predictor measurement. Results: The regression analysis shows a positive correlation between the FD‐3 and Predictor data of QRS, RMS, and LAS at both filter settings except for LAS at 25–250 Hz. Regarding the late potential classification, at 25–250 Hz 4 of 22 (18%) FD‐3 and Predictor classifications were not coherent and at 40–250 Hz only 2 of 22 (9%) classifications did not correlate. Conclusions: The FD‐3 recorder is feasible to perform a signal‐averaged ECG analysis. Especially at 40–250 Hz, the late potential classification shows a concordant distribution.
No takes yet. Share an insight, caveat, or question.
Schneider et al. (1996) reported an observational. Digital recorder FD-3 vs. Predictor bedside real-time recording system was evaluated on Correlation of QRS, RMS, and LAS mean values and late potential classification. The FD-3 digital Holter recorder showed positive correlation with the Predictor system for signal-averaged ECG, with 91% concordance in late potential classification at 40-250 Hz.