Key result
The Heptest POC-Hi whole blood assay correlated strongly with anti-Xa plasma levels during cardiopulmonary bypass (r=0.57, p<0.0001), whereas activated clotting time showed no significant relationship.
Why the study?
Does the Heptest POC-Hi assay accurately measure heparin anticoagulant activity compared to ACT and anti-Xa levels in patients undergoing CPB surgery?
Observational (n=125)
No
Does the Heptest POC-Hi assay accurately measure heparin anticoagulant activity compared to ACT and anti-Xa levels in patients undergoing CPB surgery?
Effect estimate: r=0.87
p-value: p=<0.0001
The HPOCH assay correlates strongly with anti-Xa plasma levels during CPB and is not influenced by factors affecting ACT, making it a promising tool for specific on-site measurement of heparin activities.
HPOCH may support heparin monitoring over ACT during CPB; Level 5 data leaves open confirmation before practice change.
The activated clotting time (ACT) generally used for monitoring heparinization during cardiopulmonary bypass (CPB) surgery does not specifically measure heparin anticoagulant activities. This may result in heparin over- or under-dose and subsequent severe adverse events. A new point-of-care whole blood clotting assay (Heptest POC-Hi [HPOCH]) for quantifying heparin anticoagulant activity specifically was compared with ACT and anti-factor Xa (anti-Xa) heparin plasma levels (Coatest heparin) in 125 patients undergoing CPB surgery. The analytical reliability of the HPOCH and the influence of preanalytical variables on assay results were also examined. The ACT and HPOCH clotting times determined throughout the entire observation period correlated closely (n=683; r = 0.80; p < .0001). Similarly, there was a significant linear correlation between HPOCH and Coatest anti-Xa levels (n=352; r = 0.87; p < .0001). Pre- and post-CBP values of HPOCH, ACT, and anti-Xa plasma levels correlated closely with each other (correlation coefficients between r = 0.90 and r = 0.99; p < .0001). During CPB, there was no significant relationship between ACT and whole blood or plasma heparin levels determined by HPOCH (n=157; r = 0.19) and the chromogenic anti-Xa assay (n=157; r = 0.04), respectively. In contrast, HPOCH and anti-Xa plasma levels correlated strongly during CPB (n=157; r = 0.57; p < .0001). However, bias analysis showed that the HPOCH and Coatest heparin could not be used interchangeably. The HPOCH was well reproducible and not influenced by aprotinin, hemodilution, or other factors affecting ACT. The HPOCH seems to be a promising new tool for specific on-site measurement of heparin activities in whole blood during CPB.
No takes yet. Share an insight, caveat, or question.
Hellstern et al. (2007) conducted an observational in Cardiopulmonary bypass surgery (n=125). Heptest POC-Hi (HPOCH) assay vs. Activated clotting time (ACT) and Coatest heparin (anti-Xa) was evaluated on Correlation between HPOCH and Coatest anti-Xa levels (r=0.87, p=<0.0001). The Heptest POC-Hi whole blood assay correlated strongly with anti-Xa plasma levels during cardiopulmonary bypass (r=0.57, p<0.0001), whereas activated clotting time showed no significant relationship.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: